Device recall: GS70 Salus Surgical Table, Product Code 18-070-70

Class II FDA enforcement event 92290. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Skytron, LLC started this device recall on 1 Aug 2023. The firm address in the FDA record is in Grand Rapids, MI. FDA put the recall in Class II. FDA event 92290 has 2 product records. The first FDA enforcement report date is 6 Sep 2023. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide - US Nationwide distribution in the states of AL, AR, CA, CT, FL, GA, IA, IL, IN, MA, MI, MO, PA, TN, TX, UT, WA, WI and the countries of Australia, New Zealand.

Recalling firm

Skytron, LLC, Grand Rapids, MI.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 92290: product records
Recall numberProductQuantityClassStatusReport date
Z-2469-2023GS70 Salus Surgical Table, Product Code 18-070-70110 unitsClass IIOngoing6 Sep 2023
Z-2470-2023GS70H Salus Surgical Table, Product Code 18-071-7024 unitsClass IIOngoing6 Sep 2023

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2023
917
Share of these events in Class II
88%
Product records in this event
2
Events of the firm on the register
1

Official channels

Cite this page

Recall Register. FDA enforcement event 92290. https://recallregister.com/fda/ii/92290/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide - US Nationwide distribution in the states of AL, AR, CA, CT, FL, GA, IA, IL, IN, MA, MI, MO, PA, TN, TX, UT, WA, WI and the countries of Australia, New Zealand.
What quantity of product is in FDA event 92290?
The event has 2 product records. The first record gives this quantity: 110 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2023 have the same class?
88% of the 917 device events that started in 2023 are in Class II.