Device recall: Input Output Module (IOM) FLX-201-Laboratory Automation system Brand names of the same product: FlexLab (FLX)…

Class II FDA enforcement event 91950. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus,…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

The recalling firm started this device recall on 10 Mar 2023. The FDA record names a person as the firm, and this site does not show the name. The firm address in the FDA record is in Lugano. FDA put the recall in Class II. FDA event 91950 has 5 product records. The first FDA enforcement report date is 10 May 2023. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, DENMARK, EAU, EGYPT, FRANCE, GERMANY,…

Recalling firm

The name is withheld, Lugano, Switzerland.

The FDA record gives the name of a person as the recalling firm. This site does not show the names of persons.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

5 product records in the event.

FDA event 91950: product records
Recall numberProductQuantityClassStatusReport date
Z-1559-2023Input Output Module (IOM) FLX-201-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) Part Numbers: FLX-201-00, APT-201-00, APT-253-00 ACP-201-00, FLX-201-10, FLX-253-10, FLX-253-11US:245 (16 FlexLab systems, 121 Accelerator a3600 system, 108 Aptio Automation systems).Class IIOngoing10 May 2023
Z-1560-2023Storage and Retrieval Module (SRM)-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) Part Numbers: ACP-207-00, ACP-207-01, ACP-230-00, ACP-230-01 APT-207-00, APT-207-01, APT-230-00, APT-230-01 FLX-207-00, FLX-207-01, FLX-207-02, FLX-230-00,…US: 245 (16 FlexLab systems, 121 Accelerator a3600 system, 108 Aptio Automation systems).Class IIOngoing10 May 2023
Z-1561-2023PVT Interface Module (PVT)-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) Part Numbers: 72747000.AUS: 245 (16 FlexLab systems, 121 Accelerator a3600 system, 108 Aptio Automation systems).Class IIOngoing10 May 2023
Z-1562-2023Vesmatic Cube 80 Interface Module (VMC)-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) Part Numbers: 72747000.AUS: 245 (16 FlexLab systems, 121 Accelerator a3600 system, 108 Aptio Automation systems).Class IIOngoing10 May 2023
Z-1563-2023Alinity h Interface Module (HSQ)-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) Part Numbers: FLX-274-20US:245 (16 FlexLab systems, 121 Accelerator a3600 system, 108 Aptio Automation systems).Class IIOngoing10 May 2023

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2023
917
Share of these events in Class II
88%
Product records in this event
5

Official channels

Cite this page

Recall Register. FDA enforcement event 91950. https://recallregister.com/fda/ii/91950/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, DENMARK, EAU, EGYPT, FRANCE, GERMANY,…
What quantity of product is in FDA event 91950?
The event has 5 product records. The first record gives this quantity: US:245 (16 FlexLab systems, 121 Accelerator a3600 system, 108 Aptio Automation systems).
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2023 have the same class?
88% of the 917 device events that started in 2023 are in Class II.