Recall Register

Food, drugs, devices › Stryker Sustainability Solutions

Device recall: Color Cuff Single Port, Single Bladder 30 (Royal Blue) Quick Connect, REF: 5921-030-135; Color Cuff Dual Port, Single Bl

FDA enforcement event 91845 · Class II · initiated 1 Mar 2023 · status Ongoing

DeviceVoluntary: Firm initiated1 product recordTempe, AZ

Stryker Sustainability Solutions (Tempe, AZ) initiated this device recall on 1 Mar 2023; FDA lists 1 product record under event 91845, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 88% of the 917 device enforcement events initiated in 2023 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Stryker Sustainability Solutions has 36 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Stryker Sustainability Solutions, Tempe, AZ.

Reason for recall

"Increased incidence of disposable tourniquet cuffs "not able to achieve desired pressure" and "not holding pressure" due to 1) Cuff fabric tearing during use, and 2) Pinched or unseated O-ring at the connection between the cuff, disposable fill line, and/or cuff hosing, which may lead to loss of vascular occlusion, which may lead to operative site blood loss, hemorrhage, or other complications."

Distribution

"US: WI, MD, NY, PA, WA, CO, GA, FL, OH, NC, TX, NJ, OR, MT, CA, RI, SC, NE, MI, LA, OK, SD, AL, VA, MS, WV, AZ, KY, IL, AR, ME, ND, MO, KS, HI, NH, AK, TN, IA, MN, IN, NV, MA, ID, DE, NM, WY, DC,…"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1315-2023Color Cuff Single Port, Single Bladder 30 (Royal Blue) Quick Connect, REF: 5921-030-135; Color Cuff Dual Port, Single Bladder 30 (Royal Blue) Quick Connect, REF: 5921-030-235; Color Cuff Single Port, Single Bladder 34 (Purple), Quick Connect, REF: 5921-034-135; Color Cuff Dual Port, Single Bladder 34 (Purple), Quick…127,894Class IIOngoing5 Apr 2023

Context

88%of 2023 device events are Class II917 events that year
1product record in this eventfirm median: 1
36events on this register for the firm

Other enforcement events for Stryker Sustainability Solutions

InitiatedProduct (first record)TypeClassRecords
19 Aug 2026REF:2B5LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/OPTIVIEW TECHNOLOGY & SLEEVE 5MM X 100MMDeviceClass II6
10 Apr 2026BARD Dynamic Deca Steerable, Product Number 201101; REPROCESSED ELECTROPHYSIOLOGY CATHETERDeviceClass II26
12 Mar 2026Stryker Sustainability Solution Color Cuff 24" (Yellow) Quick Connect Dual Port Single Bladder REF 5921-024-235 Stryker DeviceClass II1
24 Sep 2025HARMONIC FOCUS Shears + Adaptive Tissue Technology, 9 cm Length REF: HAR9F, Reprocessed Device for Single Use, STERILEEODeviceClass II1
2 Jul 2024Disposable 5 Lead Cable and Lead Wire System, 50 inches, REF: LW-309DS50/5ADeviceClass II1
31 May 2024REF: 60-7070-005, Zimmer, Tourniquet Cuff 30" (Blue), Luer-Lock Dual Port, single Bladder, RxOnlyDeviceClass II1
10 May 2024Stryker Sustainability Solutions, BW Lasso 2515 NAV eco Variable Diagnostic EP Catheter, REF: D134301, ELECTRODES: 22, sDeviceClass II1
21 Jun 2023HoverMatt, 39" W x 78" L, REF: HM39SPU-BDeviceClass II1
14 Apr 2023REF 5065, Kendall, Impad Rigid Sole Foot Cover, Regular Pair ,<12", Rx Only Non-SterileDeviceClass II1
21 Jan 2022REF: 5921-018-135, Stryker Color Cuff, 18" (Red), Quick Connect Single Port, Single BladderDeviceClass II1

All 36 events for Stryker Sustainability Solutions

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Increased incidence of disposable tourniquet cuffs "not able to achieve desired pressure" and "not holding pressure" due to 1) Cuff fabric tearing during use, and 2) Pinched or unseated O-ring at the connection between the cuff, disposable fill line, and/or cuff hosing, which may lead to loss of vascular occlusion, which may lead to operative site blood loss, hemorrhage, or other complications."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "US: WI, MD, NY, PA, WA, CO, GA, FL, OH, NC, TX, NJ, OR, MT, CA, RI, SC, NE, MI, LA, OK, SD, AL, VA, MS, WV, AZ, KY, IL, AR, ME, ND, MO, KS, HI, NH, AK, TN, IA, MN, IN, NV, MA, ID, DE, NM, WY, DC,…"

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.