Device recall: CardioMEMS HF System PA Sensor and Delivery System, Model Number CM2000, The CardioMEMS HF System provides pulmonary…
Class II FDA enforcement event 91606. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended radiofrequency has the potential to result in inaccurate readings or sensor signal acquisition difficulties.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
St. Jude Medical started this device recall on 7 Feb 2023. The firm address in the FDA record is in Atlanta, GA. FDA put the recall in Class II. FDA event 91606 has 3 product records. The first FDA enforcement report date is 15 Mar 2023. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait,…
Recalling firm
St. Jude Medical, Atlanta, GA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
3 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1223-2023 | CardioMEMS HF System PA Sensor and Delivery System, Model Number CM2000, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure… | 36875 units | Class II | Ongoing | 15 Mar 2023 |
| Z-1224-2023 | CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3000, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure… | 1905 units | Class II | Ongoing | 15 Mar 2023 |
| Z-1225-2023 | CardioMEMS HF System PA Sensor and Delivery System, Model Number CM3100, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which physicians use to initiate or modify heart failure… | 317 units | Class II | Ongoing | 15 Mar 2023 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2023
- 917
- Share of these events in Class II
- 88%
- Product records in this event
- 3Median of the firm: 1
- Events of the firm on the register
- 15
Other events of the firm
Official channels
- FDA enforcement report Z-1223-2023The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 91606. https://recallregister.com/fda/ii/91606/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait,…
- What quantity of product is in FDA event 91606?
- The event has 3 product records. The first record gives this quantity:
36875 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2023 have the same class?