Device recall: Digital VIPR ROC-9839 Roberts Oxygen Ready to Use Digital VIPR (head only) ROC-9840, Roberts Oxygen Ready to Use…
Class II FDA enforcement event 91561. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Gauge malfunction: Digital gauge displays an incorrect tank pressure - displays pressure higher than actual tank pressure. This may lead to tank running out of oxygen during use without audible alarms, potentially leading to an out of oxygen event and no flow.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Western/Scott Fetzer Company started this device recall on 3 Jan 2023. The firm address in the FDA record is in Westlake, OH. FDA put the recall in Class II. FDA event 91561 has 1 product record. The first FDA enforcement report date is 19 Apr 2023. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US Nationwide distribution in the states of CO, CT, ID, MD, MT, TN, WA.
Recalling firm
Western/Scott Fetzer Company, Westlake, OH.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1376-2023 | Digital VIPR ROC-9839 Roberts Oxygen Ready to Use Digital VIPR (head only) ROC-9840, Roberts Oxygen Ready to Use Digital VIPR System (with E-size cylinder) MNDR-600, Western Oxytote DTE Digital VIPR (head only) MNDS-602, Western Oxytote DTE Digital VIPR System (with D-size cylinder) MNDS-603, Western Oxytote DTE… | 24,287 units | Class II | Ongoing | 19 Apr 2023 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2023
- 917
- Share of these events in Class II
- 88%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 5
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 16 Oct 2021 | Roberts Digital VIPR System (with E-size cylinder) Model: ROC-9840 | Device | Class II | 4 |
| 15 Aug 2019 | Grab n Go Opti series VIPR system Model # PRX-9615 - Product Usage: The Grab n Go Opti system is an integrated delivery | Device | Class II | 5 |
| 31 Jan 2015 | Various OxyTOTE Portable Oxygen System regulators, Valve Integrated Pressure Regulator, one unit per package. Product Us | Device | Class I | 1 |
| 7 Apr 2014 | Model RG3000 and RG4000 Series Valve Integrated Pressure Regulator (VIPR) | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-1376-2023The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 91561. https://recallregister.com/fda/ii/91561/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US Nationwide distribution in the states of CO, CT, ID, MD, MT, TN, WA.
- What quantity of product is in FDA event 91561?
- The FDA record gives this quantity:
24,287 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2023 have the same class?