Recall Register

Device recall: 2008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a…

Class II FDA enforcement event 91154. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

(1)Software version 2.74 upgrade and (2)hardware related cause that affects only CDX Machines, has been identified as the UI-MICs Board is sensitive to Electromagnetic Interference (EMI) which will cause the false alarm "Remove USB 2 Device".

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Fresenius Medical Care Holdings, Inc. started this device recall on 27 Jul 2022. The firm address in the FDA record is in Waltham, MA. FDA put the recall in Class II. FDA event 91154 has 1 product record. The first FDA enforcement report date is 28 Dec 2022. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide distribution.

Recalling firm

Fresenius Medical Care Holdings, Inc., Waltham, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 91154: product records
Recall numberProductQuantityClassStatusReport date
Z-0734-20232008T and 2008TBluestar Hemodialysis Machines with CDX-Indicated for acute and chronic dialysis therapy in a healthcare facility Models: (1) 190713 2008T HEMODIALYSIS SYS., with CDX; (2) 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG W/CDX ; (3) 191124 2008T HD SYS. CDX BLUESTAR; (4) 191126 2008T HD SYS. CDX W/BIBAG BLUESTAR2008T with CDX= 70,552 2008T Bluestar with CDX= 25,205Class IIOngoing28 Dec 2022

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2022
839
Share of these events in Class II
88%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
66

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 91154. https://recallregister.com/fda/ii/91154/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide distribution.
What quantity of product is in FDA event 91154?
The FDA record gives this quantity: 2008T with CDX= 70,552 2008T Bluestar with CDX= 25,205
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2022 have the same class?
88% of the 839 device events that started in 2022 are in Class II.