Device recall: Product/Part: MitraClip XTR Clip Delivery System US/CDS0601-XTR, MitraClip G4 Clip Delivery System NT US/CDS0701-NT…
Class II FDA enforcement event 90840. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
The recalling firm started this device recall on 8 Sep 2022. The FDA record names a person as the firm, and this site does not show the name. The firm address in the FDA record is in Santa Clara, CA. FDA put the recall in Class II. FDA event 90840 has 3 product records. The first FDA enforcement report date is 26 Oct 2022. The record gives the status Ongoing. 92 of the 839 device events that started in 2022 have more product records. The median device event of 2022 has 1 product record. The openFDA file has 839 device events that started in 2022 and 900 that started in 2021.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide distribution - US Nationwide including in the states of PR, WA, NE, MO, IA, KS, TN, MN, ND, MD, VA, MA, PA, DC, LA, KY, CA, NV, IL, WI, TX, CT, VT, AL, AR, GA, MS, MI, OH, FL, NC, NJ, CO,…
Recalling firm
The name is withheld, Santa Clara, CA.
The FDA record gives the name of a person as the recalling firm. This site does not show the names of persons.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
3 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0115-2023 | Product/Part: MitraClip XTR Clip Delivery System US/CDS0601-XTR, MitraClip G4 Clip Delivery System NT US/CDS0701-NT, MitraClip G4 Clip Delivery System NTW US/CDS0701-NTW, MitraClip G4 Clip Delivery System XT US/CDS0701-XT, MitraClip G4 Clip Delivery System XTW US/CDS0701-XTW. For cardiac procedures. | 18,042 | Class II | Ongoing | 26 Oct 2022 |
| Z-0116-2023 | Product/Part: MitraClip NTR Clip Delivery System US/CDS0601-NTR, MitraClip NTR Clip Delivery System CE/CDS0602-NTR, MitraClip XTR Clip Delivery System CE/CDS0602-XTR, MitraClip G4 Clip Delivery System NT CE/CDS0702-NT, MitraClip G4 Clip Delivery System NTW CE/CDS0702-NTW, MitraClip G4 Clip Delivery System XT… | 20,045 | Class II | Ongoing | 26 Oct 2022 |
| Z-0117-2023 | Product/Part: MitraClip G4 CDS NT REPAIR-MR Clinical/CDS0704-NT, MitraClip G4 CDS NTW REPAIR-MR Clinical/CDS0704-NTW, MitraClip G4 CDS XT REPAIR-MR Clinical/CDS0704-XT, MitraClip G4 CDS XTW REPAIR-MR Clinical/CDS0704-XTW, TriClip NT Clip Delivery Sys CLIN/TCDS0203-NT, TriClip XT Clip Delivery Sys CLIN/TCDS0203-XT,… | 767 | Class II | Ongoing | 26 Oct 2022 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2022
- 839
- Share of these events in Class II
- 88%
- Product records in this event
- 3
Official channels
- FDA enforcement report Z-0115-2023The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 90840. https://recallregister.com/fda/ii/90840/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide distribution - US Nationwide including in the states of PR, WA, NE, MO, IA, KS, TN, MN, ND, MD, VA, MA, PA, DC, LA, KY, CA, NV, IL, WI, TX, CT, VT, AL, AR, GA, MS, MI, OH, FL, NC, NJ, CO,…
- What quantity of product is in FDA event 90840?
- The event has 3 product records. The first record gives this quantity:
18,042
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2022 have the same class?
- 88% of the 839 device events that started in 2022 are in Class II.