Device recall: GE Healthcare MR Systems, Magnetic Resonance imaging (MRI). SIGNA Premier, SIGNA Architect, SIGNA Pioneer, Discovery…
Class II FDA enforcement event 89842. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
During the de-installation of an MR system magnet, if all rigging and hardware that attach to the magnet for transportation are not properly installed and secured, the magnet may fall, leading to potential injury.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
GE Healthcare, LLC started this device recall on 4 Mar 2022. The firm address in the FDA record is in Waukesha, WI. FDA put the recall in Class II. FDA event 89842 has 1 product record. The first FDA enforcement report date is 27 Apr 2022. The record gives the status Ongoing.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide Distribution: US (nationwide), including Puerto Rico and the U.S. Virgin Islands; and OUS (Foreign) was made to Albania, Algeria, Argentina, Armenia, ASEAN HQ, Australia, Austria,…
Recalling firm
GE Healthcare, LLC, Waukesha, WI.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0983-2022 | GE Healthcare MR Systems, Magnetic Resonance imaging (MRI). SIGNA Premier, SIGNA Architect, SIGNA Pioneer, Discovery MR750w 3.0T, Discovery MR750 3.0T, SIGNA PET/MR, 1.5T SIGNA Creator, 1.5T SIGNA Explorer, 1.5T SIGNA MR355, 1.5T SIGNA MR360, 1.49T SIGNA MR380, SIGNA Voyager, SIGNA Voyager Premier Edition, SIGNA… | 18,411 systems | Class II | Ongoing | 27 Apr 2022 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2022
- 839
- Share of these events in Class II
- 88%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 225
Other events of the firm
Official channels
- FDA enforcement report Z-0983-2022The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 89842. https://recallregister.com/fda/ii/89842/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide Distribution: US (nationwide), including Puerto Rico and the U.S. Virgin Islands; and OUS (Foreign) was made to Albania, Algeria, Argentina, Armenia, ASEAN HQ, Australia, Austria,…
- What quantity of product is in FDA event 89842?
- The FDA record gives this quantity:
18,411 systems
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2022 have the same class?
- 88% of the 839 device events that started in 2022 are in Class II.