Recall Register

Device recall: Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML…

Class II FDA enforcement event 89602. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Cytocell Ltd. started this device recall on 18 Jan 2022. The firm address in the FDA record is in Cambridge. FDA put the recall in Class II. FDA event 89602 has 1 product record. The first FDA enforcement report date is 16 Mar 2022. The record gives the status Terminated, with the termination date 8 Nov 2024.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide distribution in the states of Hawaii, Missouri, Pennsylvania, Tennessee.

Recalling firm

Cytocell Ltd., Cambridge, N/A, United Kingdom.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 89602: product records
Recall numberProductQuantityClassStatusReport date
Z-0741-2022Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Number: USA-LPH02419 kitsClass IITerminated16 Mar 2022

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2022
839
Share of these events in Class II
88%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
13

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 89602. https://recallregister.com/fda/ii/89602/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide distribution in the states of Hawaii, Missouri, Pennsylvania, Tennessee.
What quantity of product is in FDA event 89602?
The FDA record gives this quantity: 19 kits
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2022 have the same class?
88% of the 839 device events that started in 2022 are in Class II.