Device recall: Nextra CH Cannulated Hammertoe Mini Instrument Kit-indicated for small bone reconstruction limited to inter-digital…
Class II FDA enforcement event 89588. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
During insertion of the middle phalanx implant into bone, the driver may deform the implant locking mechanism and result in inadequate mating force between the middle and proximal phalanx implants, resulting in surgical delay, in-situ removal of the implant, use of alternative surgical approach, disassembly of the implants post-operatively, and revision
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Nextremity Solutions started this device recall on 13 Jan 2022. The firm address in the FDA record is in Warsaw, IN. FDA put the recall in Class II. FDA event 89588 has 2 product records. The first FDA enforcement report date is 16 Mar 2022. The record gives the status Terminated, with the termination date 24 May 2024.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US Nationwide distribution in the states of AZ, IL, GA, LA, MD, NY, PA, TX, WA.
Recalling firm
Nextremity Solutions, Warsaw, IN.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
2 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0739-2022 | Nextra CH Cannulated Hammertoe Mini Instrument Kit-indicated for small bone reconstruction limited to inter-digital repair and fusion of the lesser toes SKU: CH-MINI-KT | 23 units | Class II | Terminated | 16 Mar 2022 |
| Z-0740-2022 | Nextra Cannulated Hammertoe Standard Instrument Kit-indicated for small bone reconstruction limited to inter-digital repair and fusion of the lesser toe SKU: CH-STD-KT | 24 units | Class II | Terminated | 16 Mar 2022 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2022
- 839
- Share of these events in Class II
- 88%
- Product records in this event
- 2Median of the firm: 2
- Events of the firm on the register
- 7
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 10 Feb 2022 | Nextra Hammertoe Correction System 4.5 Middle 3.2 Proximal Kit SKU: NX-4532K Orthopedic surgical instrument | Device | Class II | 5 |
| 26 Mar 2021 | InCore Lapidus Disposable Kit 28mm Left, Part Number: IC-LAP-28L - Product Usage: Intended as a three-part construct int | Device | Class II | 2 |
| 12 Feb 2020 | Nextra Hammertoe Correction System-4.5MM Middle, 3.2MM Proximal, SS, Sterile. REF: NX-4532K - Product Usage: The Nextra | Device | Class II | 1 |
| 16 Dec 2019 | Stratum Lapidus Plate - Small Rt, Part No. STRM-LAP-SMR, UDI 00817701025467. The product is part of the Stratum Foot Pla | Device | Class II | 1 |
| 11 Nov 2019 | Stratum MDS Screw 3.5x46mm ST Part Number: STRM-MDS-3546ST | Device | Class II | 3 |
| 9 Oct 2019 | Stratum Foot Plating System-NL LP Screw 3.5x50mm ST, sterile. Catalog Number: STRM-NL-3550ST - Product Usage: The Stratu | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-0739-2022The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 89588. https://recallregister.com/fda/ii/89588/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US Nationwide distribution in the states of AZ, IL, GA, LA, MD, NY, PA, TX, WA.
- What quantity of product is in FDA event 89588?
- The event has 2 product records. The first record gives this quantity:
23 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2022 have the same class?