Recall Register

Food, drugs, devices › Canon Medical System, USA, INC.

Device recall: Flat Panel Detector, Model Number: TFP-800A, used with the Alphenix Systems, Item Numbers: INFX-8000V, INFX-8000C, INFX-

FDA enforcement event 89218 · Class II · initiated 20 Dec 2021 · status Ongoing

DeviceVoluntary: Firm initiated1 product recordTustin, CA

FDA enforcement event 89218 is a device recall by Canon Medical System, USA, INC. of Tustin, CA, initiated 20 Dec 2021 and classified Class II, with 1 product record.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 89% of the 900 device enforcement events initiated in 2021 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Canon Medical System, USA, INC. has 27 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Canon Medical System, USA, INC., Tustin, CA.

Reason for recall

"Abnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a catheter from the patient's blood vessel and damage the blood vessel."

Distribution

"US Nationwide distribution in the states of DE, FL, LA, MT, NJ, OH, and WV."

Product records

Recall numberProductQuantityClassStatusReport date
Z-0492-2022Flat Panel Detector, Model Number: TFP-800A, used with the Alphenix Systems, Item Numbers: INFX-8000V, INFX-8000C, INFX-8000F, and INFX-8000H.8Class IIOngoing19 Jan 2022

Context

89%of 2021 device events are Class II900 events that year
1product record in this eventfirm median: 1
27events on this register for the firm

Other enforcement events for Canon Medical System, USA, INC.

InitiatedProduct (first record)TypeClassRecords
4 Aug 2026Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnDeviceClass II1
19 Dec 2025VANTAGE GALAN 3T Model MRT-3020/MEXL-3020DeviceClass II2
19 Dec 2025Alphenix INFX-8000C, interventional fluoroscopic x-ray systemDeviceClass II4
16 Oct 2025VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.DeviceClass II1
2 Sep 2025This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuratioDeviceClass II1
17 Sep 2024Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that combines PosDeviceClass II1
10 May 2024MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504), DeviceClass II1
29 Jan 2024MRI system: Vantage Orian, Vantage Fortian, Model: MRT-1550DeviceClass II3
15 Nov 2023The device is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in an interventional DeviceClass II2
22 Jun 2023INFX-8000C XGCP-930AA TABLE SIDE CONSOLE, INFX-8000C XGCP-930BA TABLE SIDE CONSOLEDeviceClass II1

All 27 events for Canon Medical System, USA, INC.

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Abnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a catheter from the patient's blood vessel and damage the blood vessel."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "US Nationwide distribution in the states of DE, FL, LA, MT, NJ, OH, and WV."

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.