Recall Register

Device recall: Atellica CH 930 Analyzer with Software version: V1.25.1 and lower- An automated, clinical chemistry analyzer designed…

Class II FDA enforcement event 89160. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

(1)Software (SW) versions V1.25.1 and lower may result in TDef (Test Definition) Parameters for Open Channel Assays reverting to Default Value would shift test results by the magnitude of the implemented Assay Correlation factor; (2) On Board Stability (OBS) Not Updating with Manual Change for Open Channel Assays may continue to use the reagent past its expiration date leading to potential…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Siemens Healthcare Diagnostics, Inc. started this device recall on 18 Nov 2021. The firm address in the FDA record is in Tarrytown, NY. FDA put the recall in Class II. FDA event 89160 has 1 product record. The first FDA enforcement report date is 12 Jan 2022. The record gives the status Ongoing. 291 of the 900 device events that started in 2021 have more product records. The median device event of 2021 has 1 product record. The openFDA file has 900 device events that started in 2021 and 916 that started in 2020.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Nationwide Foreign: AU DE AE TR BG CL IN CZ BR ES KE CN PL SA BE IQ IT CW SE AR CO GR SK EE KR JP UY DK FR GB VN CA BD MX TH DZ PT RU ZA PK HU NL VA NO AT IR LV EG PH SG MY IL FI HR MA KW CH TW BH…

Recalling firm

Siemens Healthcare Diagnostics, Inc., Tarrytown, NY.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 89160: product records
Recall numberProductQuantityClassStatusReport date
Z-0449-2022Atellica CH 930 Analyzer with Software version: V1.25.1 and lower- An automated, clinical chemistry analyzer designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 110670002,980 unitsClass IIOngoing12 Jan 2022

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2021
900
Share of these events in Class II
89%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
334

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 89160. https://recallregister.com/fda/ii/89160/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Nationwide Foreign: AU DE AE TR BG CL IN CZ BR ES KE CN PL SA BE IQ IT CW SE AR CO GR SK EE KR JP UY DK FR GB VN CA BD MX TH DZ PT RU ZA PK HU NL VA NO AT IR LV EG PH SG MY IL FI HR MA KW CH TW BH…
What quantity of product is in FDA event 89160?
The FDA record gives this quantity: 2,980 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2021 have the same class?
89% of the 900 device events that started in 2021 are in Class II.