Device recall: Invacare TDX SP2 with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Mod

FDA enforcement event 89062 · Class II · initiated 12 Nov 2021 · status Ongoing

DeviceVoluntary: Firm initiated3 product recordsElyria, OH

Invacare Corporation (Elyria, OH) initiated this device recall on 12 Nov 2021; FDA lists 3 product records under event 89062, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 89% of the 900 device enforcement events initiated in 2021 carry the same class.

Status: Ongoing. Initiation: voluntary: firm initiated. Invacare Corporation has 26 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Invacare Corporation, Elyria, OH.

Reason for recall

"Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury"

Distribution

"Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand."

Product records

Recall numberProductQuantityClassStatusReport date
Z-0445-2022Invacare TDX SP2 with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: TDX SP2: TDXSP2, TDXSP2-CG, TDXSP2X-CG, TDXSP2-MCG, TDXSP2X-MCG, TDXSP2V, TDXSP2V-HD, TDXSP2HD, TDXSP2HD-CG, TDXSP2HD-MCG1477 units US; 1226 units OUSClass IIOngoing12 Jan 2022
Z-0446-2022Invacare AVIVA FX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: IFX-20R, IFX-20SP, IFX-20MP, IFX-20C303 units US; 20 units OUSClass IIOngoing12 Jan 2022
Z-0447-2022Invacare AVIVA Storm RX with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DLX-GRY101-A) Models: SRX-20R, SRX-20SP, SRX-20MP1 unit USClass IIOngoing12 Jan 2022

Context

89%of 2021 device events are Class II900 events that year
3product records in this eventfirm median: 1
26events on this register for the firm

Other enforcement events for Invacare Corporation

InitiatedProduct (first record)TypeClassRecords
22 Jan 2024Invacare Homecare Series Bed and Invacare G-Series Beds Containing Component Model Numbers: (1) G50 (2) G53 (3) G54 (4) DeviceClass II1
15 Jun 2023DOLOMITE GLOSS ROLLATOR - intended to support during walking and may also serve as a resting seat. Model Number(s): 1654DeviceClass II1
10 Jun 2022Invacare Platinum 5NXG Oxygen Concentrator, Model Numbers P5NXG1, P5NXGC1 (Canada specific model)DeviceClass III1
30 Mar 2022Invacare TDX SP2, Model Nos. TDXSP2 TDXSP2-MCG TDXSP2VDeviceClass II1
30 Oct 2020Invacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory disorders requiring supplemental oxygenDeviceClass II1
16 Nov 2018Invacare 9805 and 9805P Hydraulic Patient Lifts, Model Nos. 9805 and 9805P The intended use of a Non-AC-Powered patient DeviceClass II1
25 Apr 2017Assist Bed Rail - 6632DeviceClass II1
2 Dec 2016Bariatric Bed - BAR600IVCDeviceClass II4
21 Jun 2016lnvacare Precise Rx Pediatric Flowmeter Model Number: IRCPF16. Packed one per carton. The lnvacare Precise Rx Pediatric DeviceClass II2
15 Mar 2016Invacare¿ Bariatric Bed w/o mattress & rails, Junction Boxes, Product Usage: The Invacare Bariatric home care bed is intDeviceClass II4

All 26 events for Invacare Corporation

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury"

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand."

Is the recall still ongoing?

FDA lists the status as Ongoing. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.