Device recall: Oxygen Sensor Model K54019-2 used in the: G185 & G210 Incubators and G603 Work Chambers Oxygen Sensors supplied for use
FDA enforcement event 88277 is a device recall by CooperSurgical, Inc. of Trumbull, CT, initiated 30 Jun 2021 and classified Class II, with 1 product record.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 89% of the 900 device enforcement events initiated in 2021 carry the same class.
Status: Terminated, terminated 21 Nov 2022. Initiation: voluntary: firm initiated. CooperSurgical, Inc. has 57 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
CooperSurgical, Inc., Trumbull, CT.
Reason for recall
"Units are displaying output readings that differ from the measurements taken using a Gas Analyzer. When the display is reading a target setting of 5%, an external meter is reading between 3% and 12% and the Incubator does not alarm. May result in loss of embryo during storage."
Distribution
"Only Foreign: Bahrain, Belgium, Bulgaria, Czech Republic Denmark, Egypt, Estonia, France, Georgia, Germany, Ghana, Great Britain, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Jordan,…"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-2307-2021 | Oxygen Sensor Model K54019-2 used in the: G185 & G210 Incubators and G603 Work Chambers Oxygen Sensors supplied for use in the following incubator and work chamber products; G185 Standard 110V, G185 SensorTech 110V, G185 Standard 230V, G185 SensorTech 230V, G210 InviCell Standard, G210 InviCell Standard US, G210… | 421 units | Class II | Terminated | 25 Aug 2021 |
Context
Other enforcement events for CooperSurgical, Inc.
All 57 events for CooperSurgical, Inc.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Units are displaying output readings that differ from the measurements taken using a Gas Analyzer. When the display is reading a target setting of 5%, an external meter is reading between 3% and 12% and the Incubator does not alarm. May result in loss of embryo during storage."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Only Foreign: Bahrain, Belgium, Bulgaria, Czech Republic Denmark, Egypt, Estonia, France, Georgia, Germany, Ghana, Great Britain, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Jordan,…"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 21 Nov 2022. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.