Recall Register

Device recall: Olympus Sterile Optical Laser Fiber, single use and reusable

Class II FDA enforcement event 87814. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Product sterility is potentially compromised.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Quanta System, S.p.A. started this device recall on 27 Apr 2021. The firm address in the FDA record is in Samarate. FDA put the recall in Class II. FDA event 87814 has 3 product records. The first FDA enforcement report date is 16 Jun 2021. The record gives the status Terminated, with the termination date 5 Sep 2023.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Nationwide distribution via 6 US distributors

Recalling firm

Quanta System, S.p.A., Samarate, N/A, Italy.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 87814: product records
Recall numberProductQuantityClassStatusReport date
Z-1842-2021Olympus Sterile Optical Laser Fiber, single use and reusable26925 unitsClass IITerminated16 Jun 2021
Z-1843-2021Quanta System Sterile Optical Laser Fiber, single use2427 unitsClass IITerminated16 Jun 2021
Z-1844-2021Quanta System Sterile Optical Laser Fiber, reusable307Class IITerminated16 Jun 2021

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2021
900
Share of these events in Class II
89%
Product records in this event
3Median of the firm: 3
Events of the firm on the register
2

Other events of the firm

Other FDA enforcement events of Quanta System, S.p.A.
InitiatedProduct (first record)TypeClassRecords
7 Nov 2012Q-Plus T Powered Laser Surgical Instrument Nd:YAG 1064 nm and 532 nm: Treatment of vascular lesions and pigmented lesionDeviceClass II3

Official channels

Cite this page

Recall Register. FDA enforcement event 87814. https://recallregister.com/fda/ii/87814/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Nationwide distribution via 6 US distributors
What quantity of product is in FDA event 87814?
The event has 3 product records. The first record gives this quantity: 26925 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2021 have the same class?
89% of the 900 device events that started in 2021 are in Class II.