Recall Register

Device recall: REF 7617400, Groshong NXT ClearVue Catheter, intermediate kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741045094 -…

Class II FDA enforcement event 87594. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Catheters contain two issues: 1. the inner diameter of the distal connector may be reduced resulting in inability to pass the catheter through the connector and 2. the catch slot on the distal connecter may be deformed resulting in the inability to properly lock

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Bard Access Systems Inc. started this device recall on 7 Apr 2021. The firm address in the FDA record is in Salt Lake City, UT. FDA put the recall in Class II. FDA event 87594 has 18 product records. The first FDA enforcement report date is 26 May 2021. The record gives the status Terminated, with the termination date 1 Jun 2023.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA,…

Recalling firm

Bard Access Systems Inc., Salt Lake City, UT.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

18 product records in the event.

FDA event 87594: product records
Recall numberProductQuantityClassStatusReport date
Z-1652-2021REF 7617400, Groshong NXT ClearVue Catheter, intermediate kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741045094 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1653-2021REF 7617405J, Groshong NXT ClearVue Catheter, Basic kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741200592 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1654-2021REF 7617408, Groshong NXT ClearVue Catheter, Full Kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741035255 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1655-2021REF 7617408D, Groshong NXT ClearVue Catheter, Maximal Barrier Kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741035262 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1656-2021REF 7655405, Groshong NXT ClearVue Catheter, Basic Kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741035279 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1657-2021REF 7655405J, Groshong NXT ClearVue Catheter, Basic Kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741200615 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1658-2021REF 7812400, Replacement Connector For 4F Groshong NXT clearVue PICC with Extension Leg, Rx Only, Sterile EO, CE 2797, UDI: (01)10801741075268 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1659-2021REF 9617405, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Basic Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741035989 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1660-2021REF 9617407, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Full Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741035996 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1661-2021REF 9617408, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Full Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741036009 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1662-2021REF 9617408D, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Maximal Barrier Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741036016 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1663-2021REF 9655405, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Basic Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741036023 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1664-2021REF 69655405, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Basic Kit, 4F, Single-Lumen, TLS, UDI: (01)10801741147521 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1665-2021REF 91660417, Groshong NXT ClearVue Catheter with Sherlock 3CG Tip Positioning System (TPS) Stylet Tip, Basic Kit, 4F, Single-Lumen, TPS, CE 2797, UDI: (01)10801741107822 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1666-2021REF CK000379, Groshong NXT ClearVue Catheter with Sherlock Tip Location System (TLS) Stylet, Full Kit, 4F, Single-Lumen, TLS, CE 2797, UDI: (01)10801741102704 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1667-2021REF CK000516, Groshong NXT ClearVue Catheter with Sherlock 3CG Tip Positioning System (TPS) Stylet, Full Kit, 4F, Single-Lumen, TPS, CE 2797, UDI: (01)10801741107891 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1668-2021REF 7617405, Groshong NXT ClearVue Catheter Basic Kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741086998 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021
Z-1669-2021REF 7617407, Groshong NXT ClearVue Catheter, Full Kit, 4F, Single-Lumen, 60 cm, UDI: (01)10801741035248 - Product Usage: is designed for use when central venous catheterization is prescribed.N/AClass IITerminated26 May 2021

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2021
900
Share of these events in Class II
89%
Product records in this event
18Median of the firm: 1
Events of the firm on the register
10

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 87594. https://recallregister.com/fda/ii/87594/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide distribution in the states of AK, AL, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA,…
What quantity of product is in FDA event 87594?
The event has 18 product records. The first record gives this quantity: N/A
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2021 have the same class?