Device recall: Ultrasonic pulsed doppler imaging system - Product Usage: intended for use by trained personnel (doctor, Sonographer…

Class II FDA enforcement event 87493. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Arietta 850 software version 4.0.0., 4.0.1, and 4.0.2 has an error in the focus point and transducer aperture settings in the SWE function. When the affected software with SWE function is used in conjunction with the C252 probe, this error can result in out of specification MI/TI acoustic output. The MI/TI acoustic output is higher than regulatory limits.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Hitachi Medical Systems America Inc started this device recall on 1 Mar 2021. The firm address in the FDA record is in Twinsburg, OH. FDA put the recall in Class II. FDA event 87493 has 1 product record. The first FDA enforcement report date is 31 Mar 2021. The record gives the status Terminated, with the termination date 4 Nov 2021.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide distribution in the state of TX.

Recalling firm

Hitachi Medical Systems America Inc, Twinsburg, OH.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 87493: product records
Recall numberProductQuantityClassStatusReport date
Z-1282-2021Ultrasonic pulsed doppler imaging system - Product Usage: intended for use by trained personnel (doctor, Sonographer, etc.) for the diagnostic ultrasound evaluation of Fetal, Abdominal, Intra-operative (Spec.), Intraoperative (Neuro.), Laparoscopic, Pediatric, Small Organ (Spec.), Neonatal Cephalic, Adult Cephalic,…1Class IITerminated31 Mar 2021

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2021
900
Share of these events in Class II
89%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
15

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 87493. https://recallregister.com/fda/ii/87493/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide distribution in the state of TX.
What quantity of product is in FDA event 87493?
The FDA record gives this quantity: 1
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2021 have the same class?
89% of the 900 device events that started in 2021 are in Class II.