Device recall: Medtronic, REF: MMT-7008A, Enlite Glucose Sensor for the following packaging: ENLITE PLUS 5PK 23L 1CL /GTIN…
Class II FDA enforcement event 87476. Recall start date: . Status in the FDA record: Ongoing.
Reason for recall, in the words of FDA
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Medtronic Inc. started this device recall on 26 Feb 2021. The firm address in the FDA record is in Northridge, CA. FDA put the recall in Class II. FDA event 87476 has 20 product records. The first FDA enforcement report date is 26 May 2021. The record gives the status Ongoing. 12 of the 900 device events that started in 2021 have more product records. The median device event of 2021 has 1 product record. The openFDA file has 900 device events that started in 2021 and 916 that started in 2020.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide distribution.
Recalling firm
Medtronic Inc., Northridge, CA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
20 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1599-2021 | Medtronic, REF: MMT-7008A, Enlite Glucose Sensor for the following packaging: ENLITE PLUS 5PK 23L 1CL /GTIN: 20643169541706, ENLT PLUS 5PK 2L 1CL PR/ GTIN: 20643169786725, ENLITE PLUS 5PK 23L 1CL/ GTIN: 20763000421292 - Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to… | 213,110 units | Class II | Ongoing | 26 May 2021 |
| Z-1600-2021 | Medtronic REF: MMT-7008B, Enlite Glucose Sensor for the following packaging: ENLITE PLUS 23L 1CL / GTIN: 00643169541719 ENLITE PLUS 2L 1CL PR/ GTIN: 00643169786738 ENLITE PLUS 1PK 23L 1CL/ GTIN: 00763000421304 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously… | 20,913 units | Class II | Ongoing | 26 May 2021 |
| Z-1601-2021 | Medtronic REF: MMT-7020A, Guardian Sensor (3) for the following packaging: GS3 5PK 2L US/ GTIN: 20763000140940 GS3 5PK 2L US/ GTIN: 20763000337005 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. | 1,224,032 units | Class II | Ongoing | 26 May 2021 |
| Z-1602-2021 | Medtronic REF: MMT-7020B, Guardian Sensor (3) for the following packaging: GS3 1PK 2L US/GTIN: 00763000140779 GS3 1PK 2L US/ GTIN: 00763000337049 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. | 302,667 units | Class II | Ongoing | 26 May 2021 |
| Z-1603-2021 | Medtronic REF: MMT-7020GA, Guardian Sensor (3) for the following packaging: GS3 5PK US GC/ GTIN: 20763000179629 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. | 1,863 units | Class II | Ongoing | 26 May 2021 |
| Z-1604-2021 | Medtronic REF: MMT-7020LA, Guardian Sensor (3) for the following packaging: GS3 5PK US/ GTIN: 20763000336985 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. | 37,228 units | Class II | Ongoing | 26 May 2021 |
| Z-1605-2021 | Medtronic REF: MMT-7020LB, Guardian Sensor (3) for the following packaging: GS3 1PK US/ GTIN: 763000336998 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. | 5,571 units | Class II | Ongoing | 26 May 2021 |
| Z-1606-2021 | Medtronic REF: MMT-7020C Guardian Sensor (3) kit for the following packaging: GS3 SEMI 5PK OUS/ GTIN: 20763000192086, 5PK SF NEW GS3 OUS/GTIN: 20763000358451 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.… | 497,747 units | Class II | Ongoing | 26 May 2021 |
| Z-1607-2021 | Medtronic REF: MMT-7020C1 Guardian Sensor (3) for the following packaging: GS3 5PK OUS1/ GTIN: 20643169704071 GS3 5PK 9L OUS1/ GTIN: 20763000243320 GS3 5PK 9L OUS1/ GTIN: 20763000318844 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels… | 220,661 units | Class II | Ongoing | 26 May 2021 |
| Z-1608-2021 | Medtronic REF: MMT-7020C2 Guardian Sensor (3) for the following packaging: GS3 5PK 9L OUS2/ GTIN: 20763000243344 GS3 5PK 9L OUS2/ GTIN: 20763000318868 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product… | 52,196 units | Class II | Ongoing | 26 May 2021 |
| Z-1609-2021 | Medtronic REF: MMT-7020C3 Guardian Sensor (3) for the following packaging: GS3 5PK OUS3/ GTIN: 20643169704118 GS3 5PK 9L OUS3/ GTIN: 20763000243368 GS3 5PK 9L OUS3/ GTIN: 20763000318882 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels… | 15,393 units | Class II | Ongoing | 26 May 2021 |
| Z-1610-2021 | Medtronic REF: MMT-7020C4 Guardian Sensor (3) for the following packaging: GS3 5PK 2L CANADA/ GTIN: 20763000179612 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US. | 54,211 units | Class II | Ongoing | 26 May 2021 |
| Z-1611-2021 | Medtronic REF: MMT-7020C5 Guardian Sensor (3) for the following packaging: GS3 5PK 1L AUS/ GTIN: 20763000252056 GS3 5PK 1L AUS/ GTIN: 20763000414362 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. Product… | 68,069 units | Class II | Ongoing | 26 May 2021 |
| Z-1612-2021 | Medtronic REF: MMT-7020D1 Guardian Sensor (3) for the following packaging: GS3 1PK OUS1/ GTIN: 00643169704084 GS3 1PK 9L OUS1/ GTIN: 00763000243333 GS3 1PK 9L OUS1/ GTIN: 00763000318857 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels… | 111,678 units | Class II | Ongoing | 26 May 2021 |
| Z-1613-2021 | Medtronic REF: MMT-7020D2 Guardian Sensor (3) for the following packaging: GS3 1PK 9L OUS2/ GTIN: 00763000243357 GS3 1PK 9L OUS2/ GTIN: 00763000318871 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product… | 15,162 units | Class II | Ongoing | 26 May 2021 |
| Z-1614-2021 | Medtronic REF: MMT-7020D3 Guardian Sensor (3) for the following packaging: GS3 1 PK OUS3/ GTIN: 00643169704121 GS3 1PK 9L OUS3/ GTIN: 00763000318895 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not… | 13,014 units | Class II | Ongoing | 26 May 2021 |
| Z-1615-2021 | Medtronic REF: MMT-7020D4 Guardian Sensor (3) for the following packaging: GS3 1PK 2L CANADA/ GTIN: 00763000179595¿ Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US. | 12,624 units | Class II | Ongoing | 26 May 2021 |
| Z-1616-2021 | Medtronic REF: MMT-7020D5 Guardian Sensor (3) for the following packaging: GS3 1PK 1L AUS/ GTIN 00763000252069 GS3 1PK 1L AUS/ GTIN 00763000414375 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not… | 15,926 units | Class II | Ongoing | 26 May 2021 |
| Z-1617-2021 | Medtronic, REF: MMT-7008C, Enlite Glucose Sensor for the following packaging: ENLITE AUS 1 PRESS 5PK/ GTIN: 20763000252032 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US. | 1,174 units | Class II | Ongoing | 26 May 2021 |
| Z-1618-2021 | Medtronic, REF: MMT-7008D, Enlite Glucose Sensor for the following packaging: ENLITE AUS 1 PRESS 1PK; GTIN: 00763000252045 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US. | 207 units | Class II | Ongoing | 26 May 2021 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2021
- 900
- Share of these events in Class II
- 89%
- Product records in this event
- 20Median of the firm: 1
- Events of the firm on the register
- 49
Other events of the firm
Official channels
- FDA enforcement report Z-1599-2021The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 87476. https://recallregister.com/fda/ii/87476/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 10 Nov 2026
- Newest FDA report date on the register
- 4 Nov 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide distribution.
- What quantity of product is in FDA event 87476?
- The event has 20 product records. The first record gives this quantity:
213,110 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2021 have the same class?
- 89% of the 900 device events that started in 2021 are in Class II.