Recall Register

Food, drugs, devices › Biocare Medical, LLC

Device recall: BIOCARE Medical , IntelliPath FLX, Automated Staining Instrument, Ref: " Model Number:IPS0001US and IPS0001INTL containi

FDA enforcement event 87426 · Class II · initiated 25 Feb 2021 · status Terminated · terminated 31 Jan 2023

DeviceVoluntary: Firm initiated1 product recordPacheco, CA

Biocare Medical, LLC (Pacheco, CA) initiated this device recall on 25 Feb 2021; FDA lists 1 product record under event 87426, classified Class II.

FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 89% of the 900 device enforcement events initiated in 2021 carry the same class.

Status: Terminated, terminated 31 Jan 2023. Initiation: voluntary: firm initiated. Biocare Medical, LLC has 10 enforcement events on this register, with a median of 1 product record per event.

Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.

Recalling firm

Biocare Medical, LLC, Pacheco, CA.

Reason for recall

"Their is a potential that the automated staining instrument with software version 3.5.3.1 may move with random speeds in random direction across its range of motion. This could result in the instrument hitting end stops or running into other components on the working deck. This could also result in wash buffer being dispensed on random locations."

Distribution

"US: AL, AZ, CA, CT, FL, IL, MA, MD, MN, NC, OH, OR, PA, WA, WY OUS: Germany, Netherlands"

Product records

Recall numberProductQuantityClassStatusReport date
Z-1333-2021BIOCARE Medical , IntelliPath FLX, Automated Staining Instrument, Ref: " Model Number:IPS0001US and IPS0001INTL containing software version 3.5.3.172 unitsClass IITerminated7 Apr 2021

Context

89%of 2021 device events are Class II900 events that year
1product record in this eventfirm median: 1
10events on this register for the firm

Other enforcement events for Biocare Medical, LLC

InitiatedProduct (first record)TypeClassRecords
25 Sep 2024intelliPATH Universal HRP Detection Kit, REF: IPK5011G80, containing DAB Buffer, Catalog Number: IPBF5009G20; ONCORE ProDeviceClass II1
10 Oct 20194plus Streptavidin HRP Label, Conjugated Streptavidin Horseradish Peroxidase, Detection Component, Catalog Number: HP604DeviceClass II1
24 Apr 2018Vulcan Fast Red Chromogen Kit 2, part of an immunohistochemistry (IHC) procedure using an alkaline phosphatase (AP) deteDeviceClass II1
15 Mar 2018URO-3 Triple Stain (CD44 + p53) with CK20, Catalog #: PM370TSAA, Package Type: Bag, which contains components: a) CD44 +DeviceClass II1
31 Mar 2016Mach 4 Mouse Probe vial within the Mach 4 Universal HRP-Polymer Kit; Catalog number M4U534L; For In Vitro Use ImmunologyDeviceClass III1
8 Sep 2015Smooth Muscle Action (SMA) Immunohistochemistry reagent. Catalog numbers: CME305A-0.1ml; CME305B - 0.5 ml, PME305AA - 6.DeviceClass II1
6 May 2015D2-40 (Lymphatic Marker); For In Vitro Diagnostic; Catalog number: PM266AA Product Usage: In Vitro Diagnostic Use.: a moDeviceClass III1
17 Mar 2015CD99 Catalog number PM008 AA; For In Vitro Diagnostic Use. Product Usage: Provides adjunctive diagnostic information folDeviceClass II1
12 Dec 2014MUM-1 (Multiple myeloma oncogene-1), catalog number PRM352 AA; Product Usage: For In Vitro Diagnostic Use. This antibodyDeviceClass II1

All 10 events for Biocare Medical, LLC

Questions and answers

Why was this device recalled?

FDA's record gives the reason for recall as: "Their is a potential that the automated staining instrument with software version 3.5.3.1 may move with random speeds in random direction across its range of motion. This could result in the instrument hitting end stops or running into other components on the working deck. This could also result in wash buffer being dispensed on random locations."

What does Class II mean?

FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

Where was the product distributed?

FDA's distribution pattern field reads: "US: AL, AZ, CA, CT, FL, IL, MA, MD, MN, NC, OH, OR, PA, WA, WY OUS: Germany, Netherlands"

Is the recall still ongoing?

FDA lists the status as Terminated; the termination date is 31 Jan 2023. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.

Where does this data come from?

openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.

Provenance

Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.