Device recall: HeartSpan Transseptal Needles Catalog No. FND-019-01, Lot No. E1913644 consisting of an outer needle cannula and an…
Class II FDA enforcement event 87306. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
The labeled needle tip curvature of specific lots of transseptal needles may not match the actual needle tip curvature. In the event that the incorrect needle is unknowingly used, an unintended anatomy puncture may occur, which is likely to require medical intervention.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Merit Medical Systems, Inc. started this device recall on 19 Jan 2021. The firm address in the FDA record is in South Jordan, UT. FDA put the recall in Class II. FDA event 87306 has 2 product records. The first FDA enforcement report date is 24 Mar 2021. The record gives the status Terminated, with the termination date 17 May 2022.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
U.S. Consignees: VA, NE, Washington DC, MI, NY, KY, CA, TX, MD, AL, GA, UT, LA, CA, CT, PA, WA, MT and NJ. No government consignees. Also distributed OUS.
Recalling firm
Merit Medical Systems, Inc., South Jordan, UT.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
2 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1262-2021 | HeartSpan Transseptal Needles Catalog No. FND-019-01, Lot No. E1913644 consisting of an outer needle cannula and an inner stylet. The needle is comprised of flexible thin walled tubing with an ergonomic hub and stopcock attached to the proximal end. The stylet consists of a solid wire that when inserted in the needle… | 587 Units | Class II | Terminated | 24 Mar 2021 |
| Z-1263-2021 | Heartspan Transseptal Needles, Catalog No. FND-019-02, Lot No. E1913645 consisting of an outer needle cannula and an inner stylet. The needle is comprised of flexible thin walled tubing with an ergonomic hub and stopcock attached to the proximal end. The stylet consists of a solid wire that when inserted in the… | 448 Units | Class II | Terminated | 24 Mar 2021 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2021
- 900
- Share of these events in Class II
- 89%
- Product records in this event
- 2Median of the firm: 1
- Events of the firm on the register
- 59
Other events of the firm
Official channels
- FDA enforcement report Z-1262-2021The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 87306. https://recallregister.com/fda/ii/87306/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
U.S. Consignees: VA, NE, Washington DC, MI, NY, KY, CA, TX, MD, AL, GA, UT, LA, CA, CT, PA, WA, MT and NJ. No government consignees. Also distributed OUS.
- What quantity of product is in FDA event 87306?
- The event has 2 product records. The first record gives this quantity:
587 Units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2021 have the same class?
- 89% of the 900 device events that started in 2021 are in Class II.