Device recall: Harmony Referral System (aka Harmony RS)/ Medical Device Data System - Product Usage: intended to be used by trained…
Class II FDA enforcement event 87086. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
Harmony RS integrations with Topcon equipment, TRC NW-400 and the Signal Camera, allowed user input of non-unique patient IDs that caused multiple patient records and images to be combined under the first generated record
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Topcon Medical Systems, Inc. started this device recall on 18 Dec 2020. The firm address in the FDA record is in Oakland, NJ. FDA put the recall in Class II. FDA event 87086 has 1 product record. The first FDA enforcement report date is 10 Feb 2021. The record gives the status Terminated, with the termination date 19 Oct 2022.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
AZ, MO, NC, NY, PA, TX
Recalling firm
Topcon Medical Systems, Inc., Oakland, NJ.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0952-2021 | Harmony Referral System (aka Harmony RS)/ Medical Device Data System - Product Usage: intended to be used by trained healthcare professionals and clinicians to access archived multimedia data, forms and patient information. | 13 units | Class II | Terminated | 10 Feb 2021 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2020
- 916
- Share of these events in Class II
- 93%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 3
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 17 Sep 2024 | The IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and mana | Device | Class II | 1 |
| 6 Mar 2019 | IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, st | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-0952-2021The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 87086. https://recallregister.com/fda/ii/87086/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
AZ, MO, NC, NY, PA, TX
- What quantity of product is in FDA event 87086?
- The FDA record gives this quantity:
13 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2020 have the same class?