Device recall: GE NM/CT 850 system Model Number H3907AD - Product Usage: intended for general Nuclear Medicine imaging procedures for…

Class II FDA enforcement event 87023. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

GE Healthcare, LLC started this device recall on 10 Dec 2020. The firm address in the FDA record is in Waukesha, WI. FDA put the recall in Class II. FDA event 87023 has 6 product records. The first FDA enforcement report date is 20 Jan 2021. The record gives the status Terminated, with the termination date 21 Aug 2026.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide

Recalling firm

GE Healthcare, LLC, Waukesha, WI.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

6 product records in the event.

FDA event 87023: product records
Recall numberProductQuantityClassStatusReport date
Z-0853-2021GE NM/CT 850 system Model Number H3907AD - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.24 unitsClass IITerminated20 Jan 2021
Z-0854-2021GE NM/CT 860 system Model Number H3908AD - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.12 unitsClass IITerminated20 Jan 2021
Z-0855-2021GE NM/CT 870 CZT system Model Number H3906CW - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.12 unitsClass IITerminated20 Jan 2021
Z-0856-2021Discovery NM 630 (Model Number H3101RH), NM 830 (Model Number H3910AC), 830 NM Cyber Defense (Model Number H3910AG) - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.81 unitsClass IITerminated20 Jan 2021
Z-0857-2021Discovery 670 DR (Model H3100BT), NM/CT 870 DR (Model 3100Q and H3100AS), Discovery NM670 Pro (Model 3100XB), Discovery 670 ES (Model H3100XF) - Product Usage: intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the…52 untitsClass IITerminated20 Jan 2021
Z-0858-2021Optima NM/CT 640, Model number 3101RL - Product Usage: intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.7 unitsClass IITerminated20 Jan 2021

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2020
916
Share of these events in Class II
93%
Product records in this event
6Median of the firm: 1
Events of the firm on the register
225

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 87023. https://recallregister.com/fda/ii/87023/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide
What quantity of product is in FDA event 87023?
The event has 6 product records. The first record gives this quantity: 24 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2020 have the same class?
93% of the 916 device events that started in 2020 are in Class II.