Recall Register

Device recall: Tina-quant C-Reactive Protein Gen.3, Catalog number 05172373190 - Product Usage: the in vitro quantitative…

Class II FDA enforcement event 86745. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Roche confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the Tina-quant C-Reactive Protein Gen. 3 on the cobas c701 and 702 modules. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Roche Diagnostics Operations, Inc. started this device recall on 6 Nov 2020. The firm address in the FDA record is in Indianapolis, IN. FDA put the recall in Class II. FDA event 86745 has 1 product record. The first FDA enforcement report date is 23 Dec 2020. The record gives the status Terminated, with the termination date 18 Oct 2024.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide distribution including in the states of AL, CA, FL, GA, HI, IA, IL, IN, KY, MA, MD, MI, MN, MO, NJ, NY, OH, PA, SC, TN, TX, VA.

Recalling firm

Roche Diagnostics Operations, Inc., Indianapolis, IN.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 86745: product records
Recall numberProductQuantityClassStatusReport date
Z-0634-2021Tina-quant C-Reactive Protein Gen.3, Catalog number 05172373190 - Product Usage: the in vitro quantitative determination of CRP in human serum and plasma on Roche automated clinical chemistry analyzers.699 unitsClass IITerminated23 Dec 2020

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2020
916
Share of these events in Class II
93%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
69

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 86745. https://recallregister.com/fda/ii/86745/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide distribution including in the states of AL, CA, FL, GA, HI, IA, IL, IN, KY, MA, MD, MI, MN, MO, NJ, NY, OH, PA, SC, TN, TX, VA.
What quantity of product is in FDA event 86745?
The FDA record gives this quantity: 699 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2020 have the same class?
93% of the 916 device events that started in 2020 are in Class II.