Recall Register

Device recall: Puritan Bennett 980 (PB980) Ventilator Series - Product Usage: for use on patient population sizes from Neonatal…

Class II FDA enforcement event 86643. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

In-house lab observations and customer reports of overheating and/or thermal damage to the battery pack, create the risk of exposure to smoke or hazardous materials, which may result in burns and/or bronchospasm,

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Covidien Llc started this device recall on 16 Oct 2020. The firm address in the FDA record is in Mansfield, MA. FDA put the recall in Class II. FDA event 86643 has 1 product record. The first FDA enforcement report date is 18 Nov 2020. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide distribution and the countries of Argentina, Bolivia, Brazil, Cayman Islands, Chile, Columbia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala,…

Recalling firm

Covidien Llc, Mansfield, MA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 86643: product records
Recall numberProductQuantityClassStatusReport date
Z-0425-2021Puritan Bennett 980 (PB980) Ventilator Series - Product Usage: for use on patient population sizes from Neonatal (NICU) through Adult who require respiratory support or mechanical ventilation SKU: 980X1ENDIUU 980X1ENDIUU 980X3FRAIQC 980X3DEDRAC 980X3ENNISB 980X3NLDRAC 980X3PLDIPC 980X3ESDIEC 980X3ITDIEC 980X3JADIJJ…23,329 units total:. US 12,956; OUS 10,405Class IIOngoing18 Nov 2020

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2020
916
Share of these events in Class II
93%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
77

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 86643. https://recallregister.com/fda/ii/86643/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide distribution and the countries of Argentina, Bolivia, Brazil, Cayman Islands, Chile, Columbia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala,…
What quantity of product is in FDA event 86643?
The FDA record gives this quantity: 23,329 units total:. US 12,956; OUS 10,405
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2020 have the same class?
93% of the 916 device events that started in 2020 are in Class II.