Recall Register

Device recall: Bien-Air Surgery REF 1600407-001 SN **** FOOTCTRL OSSEODOC - Product Usage: intended to convert electrical energy to…

Class II FDA enforcement event 86419. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Bien-Air Surgery Sa started this device recall on 11 Sep 2020. The firm address in the FDA record is in Le Noirmont. FDA put the recall in Class II. FDA event 86419 has 5 product records. The first FDA enforcement report date is 28 Oct 2020. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide distribution including in the states of CA, MI and MN.

Recalling firm

Bien-Air Surgery Sa, Le Noirmont, N/A, Switzerland.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

5 product records in the event.

FDA event 86419: product records
Recall numberProductQuantityClassStatusReport date
Z-0176-2021Bien-Air Surgery REF 1600407-001 SN **** FOOTCTRL OSSEODOC - Product Usage: intended to convert electrical energy to mechanically drive handpieces and microsaws used to cut and shape bones.Total US = 1120 foot pedalsClass IIOngoing28 Oct 2020
Z-0177-2021Bien-Air Surgery REF 1600517-001 SN ******* MULTIFUNCTION PEDAL OSSEODUO - Product Usage: Product Usage: intended to convert electrical energy to mechanically drive attachment, microsaws, shaver blades and burs, burs and perforator tools for cutting and shaping bone including spine and cranium.Total US = 1120 foot pedalsClass IIOngoing28 Oct 2020
Z-0178-2021Bien-Air Surgery REF 1600686-001 SN ******* OSSEOSTAP Footcontrol - Product Usage: an active medical device that convert electrical energy into mechanical rotation to a but during Ear, Nose and Throat (ENT) surgery.Total US = 1120 foot pedalsClass IIOngoing28 Oct 2020
Z-0179-2021Stryker Instruments REF 5452-007-000 SN ******* S2 MINI FOOT Pedal - Product Usage: The S2 Mini Drill System has been designed for the light drilling of bones as part of surgical operations such as stapedotomy or ossiculoplasty. The S2 Mini Drill System consists of a foot pedal and a drill, with an integrated…Total US = 1120 foot pedalsClass IIOngoing28 Oct 2020
Z-0180-2021Entellus Medical REF SSA-102 SN ******* Shaver System Foot Pedal - Product Usage: The Entellus Medical Shaver System has been designed for shaping bone and for the resection of soft and hard tissues as part of surgical procedures in the areas of otorhinolaryngology.N/AClass IIOngoing28 Oct 2020

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2020
916
Share of these events in Class II
93%
Product records in this event
5
Events of the firm on the register
1

Official channels

Cite this page

Recall Register. FDA enforcement event 86419. https://recallregister.com/fda/ii/86419/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide distribution including in the states of CA, MI and MN.
What quantity of product is in FDA event 86419?
The event has 5 product records. The first record gives this quantity: Total US = 1120 foot pedals
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2020 have the same class?
93% of the 916 device events that started in 2020 are in Class II.