Recall Register

Drug recall: Metformin Hydrochloride Extended-Release Tablets USP, 750 mg, a) 100 ct. bottle (NDC 70010-492-01) and 500 ct. bottle…

Class II FDA enforcement event 85986. Recall start date: . Status in the FDA record: Ongoing.

Reason for recall, in the words of FDA

CGMP Deviations: FDA analysis detected N-Nitrosodimethylamine (NDMA) impurity above the acceptable intake level

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Granules Pharmaceuticals Inc started this drug recall on 3 Jul 2020. The firm address in the FDA record is in Chantilly, VA. FDA put the recall in Class II. FDA event 85986 has 1 product record. The first FDA enforcement report date is 22 Jul 2020. The record gives the status Ongoing.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide

Recalling firm

Granules Pharmaceuticals Inc, Chantilly, VA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 85986: product records
Recall numberProductQuantityClassStatusReport date
D-1386-2020Metformin Hydrochloride Extended-Release Tablets USP, 750 mg, a) 100 ct. bottle (NDC 70010-492-01) and 500 ct. bottle (NDC 70010-492-05), Manufactured for: Granules Pharmaceutical Inc., Chantilly, VA, Manufactured by: Granules India Limited, Hyderabad - 500-081, Indiaa) 476,073 bottles; b) 2,100 bottlesClass IIOngoing22 Jul 2020

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Drug events that started in 2020
342
Share of these events in Class II
64%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
7

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 85986. https://recallregister.com/fda/ii/85986/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA drug enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide
What quantity of product is in FDA event 85986?
The FDA record gives this quantity: a) 476,073 bottles; b) 2,100 bottles
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many drug recalls of 2020 have the same class?
64% of the 342 drug events that started in 2020 are in Class II.