Device recall: Dover 100% Silicone Foley Catheter with Temperature Sensor 14FR 5CC2W, Product Code: 50514T - Product Usage: for…
Class II FDA enforcement event 85629. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2°C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Covidien Llc started this device recall on 30 Apr 2020. The firm address in the FDA record is in Mansfield, MA. FDA put the recall in Class II. FDA event 85629 has 11 product records. The first FDA enforcement report date is 20 May 2020. The record gives the status Terminated, with the termination date 14 Dec 2023.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US Nationwide distribution.
Recalling firm
Covidien Llc, Mansfield, MA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
11 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1984-2020 | Dover 100% Silicone Foley Catheter with Temperature Sensor 14FR 5CC2W, Product Code: 50514T - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only. | 468 units | Class II | Terminated | 20 May 2020 |
| Z-1985-2020 | Dover 100% Silicone Silver-Hydrogel Coated Foley Catheter with Temperature Sensor 16FR 5CCCC2W, Product Code: 50516ICT - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only. | 480 units | Class II | Terminated | 20 May 2020 |
| Z-1986-2020 | Dover 100% Silicone Foley Catheter with Temperature 16FR 5CC2W, Product Code: 50516T - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only. | 3012 units | Class II | Terminated | 20 May 2020 |
| Z-1987-2020 | Dover 100% Silicone Silver-Hydrogel Coated Foley Catheter with Temperature Sensor 18FR 5CCCC2W, Product Code: 50518ICT - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only. | 0 | Class II | Terminated | 20 May 2020 |
| Z-1988-2020 | Dover 100% Silicone Premium Urine Meter Foley Tray, 400mL, 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P14TSD - Product Usage: for Drainage of the Urinary bladder and simultaneous… | 3470 units | Class II | Terminated | 20 May 2020 |
| Z-1989-2020 | Dover 100% Silicone Premium Urine Meter Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Foley Catheter Pre-connected to 2000 mL Drainage Bag with Temperature Sensor, Sterile, Product Code: P4P16TS - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only. | 8162 units | Class II | Terminated | 20 May 2020 |
| Z-1990-2020 | Dover 400mL, 100% Silicone Premium Urine Meter Foley Tray, 16 Fr/Ch (5.3 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P16TSD - Product Usage: for Drainage of the Urinary bladder and simultaneous… | 23587 units | Class II | Terminated | 20 May 2020 |
| Z-1991-2020 | Dover 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray 400mL, Premium, Urine Meter, 16 Fr/Ch (5.3 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code P4P16XTSD - Product Usage: for Drainage of… | 46828 units | Class II | Terminated | 20 May 2020 |
| Z-1992-2020 | Dover 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P16XTSD - Product Usage: for Drainage of the Urinary bladder… | 46828 units | Class II | Terminated | 20 May 2020 |
| Z-1993-2020 | Dover 400 ml 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray,18 Fr/Ch (6.0 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Temperature Sensor, Sterile, Product Code: P4P18XTS - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription… | 120 units | Class II | Terminated | 20 May 2020 |
| Z-1994-2020 | Dover 400 ml 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray, 18 Fr/Ch (6.0 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P18XTSD - Product Usage: for Drainage of the Urinary bladder… | 0 | Class II | Terminated | 20 May 2020 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2020
- 916
- Share of these events in Class II
- 93%
- Product records in this event
- 11Median of the firm: 1
- Events of the firm on the register
- 77
Other events of the firm
Official channels
- FDA enforcement report Z-1984-2020The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 85629. https://recallregister.com/fda/ii/85629/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US Nationwide distribution.
- What quantity of product is in FDA event 85629?
- The event has 11 product records. The first record gives this quantity:
468 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2020 have the same class?
- 93% of the 916 device events that started in 2020 are in Class II.