Recall Register

Device recall: DFINE Vertebral Hydraulic Assemblies: Hydraulic Assembly, StabiliT Vertebral Augmentation System, Reference/Catalog…

Class II FDA enforcement event 85485. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Vertebral hydraulic assemblies has incorrect expiration date printed on the labels.The product's shelf-life is two (2) years, but the expiration date printed on the labels is five (5) years.The affected product will begin expiring in March 2020. Using product beyond the validated shelf-life may result in (1) product not functioning as intended and/or (2) a delay in procedure or the need for the…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Merit Medical Systems, Inc. started this device recall on 13 Feb 2020. The firm address in the FDA record is in South Jordan, UT. FDA put the recall in Class II. FDA event 85485 has 2 product records. The first FDA enforcement report date is 27 May 2020. The record gives the status Terminated, with the termination date 26 Feb 2021.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution: US Nationwide distribution including in the states of (TX, UT, PA, KY, TX, AR, CA, and PA) which included one (1) government/military consignee in KY; and the countries of…

Recalling firm

Merit Medical Systems, Inc., South Jordan, UT.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 85485: product records
Recall numberProductQuantityClassStatusReport date
Z-2067-2020DFINE Vertebral Hydraulic Assemblies: Hydraulic Assembly, StabiliT Vertebral Augmentation System, Reference/Catalog number 1402 - Product Usage: is intended for percutaneous delivery of StabiliT ER2 Bone Cement in kyphoplasty procedures in the treatment of pathological fractures of the vertebrae. Painful vertebral…228Class IITerminated27 May 2020
Z-2068-2020DFINE Vertebral Assemblies: Hydraulic-Master Syringe Assembly, StabiliT Vertebral Augmentation System, Reference/Catalog number 3427 - Product Usage: is intended for percutaneous delivery of StabiliT ER2 Bone Cement in kyphoplasty procedures in the treatment of pathological fractures of the vertebrae. Painful…63Class IITerminated27 May 2020

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2020
916
Share of these events in Class II
93%
Product records in this event
2Median of the firm: 1
Events of the firm on the register
59

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 85485. https://recallregister.com/fda/ii/85485/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution: US Nationwide distribution including in the states of (TX, UT, PA, KY, TX, AR, CA, and PA) which included one (1) government/military consignee in KY; and the countries of…
What quantity of product is in FDA event 85485?
The event has 2 product records. The first record gives this quantity: 228
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2020 have the same class?
93% of the 916 device events that started in 2020 are in Class II.