Device recall: CapsoCAM Plus, UDI: 00867770000209
Class II FDA enforcement event 85378. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
Capsule Endoscopy System: a defect present in the capsule window may be stressed through the production process and in the final package, which could cause the housing of the capsule to crack and leak when ingested. Fluid from the patient could leak in to the capsule, which may damage components inside the capsule and the patient data may not be retrieved; the patient may need to repeat the exam.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Capso Vision, Inc. started this device recall on 18 Oct 2019. The firm address in the FDA record is in Saratoga, CA. FDA put the recall in Class II. FDA event 85378 has 1 product record. The first FDA enforcement report date is 6 May 2020. The record gives the status Terminated, with the termination date 8 Jan 2021.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
U.S.: CA, TX, FL, VA, PA, SC, HI, DC, NJ, UT, LA. Foreign: FR, AR, IT, UK, BE, TC, AU, CO, GM, GR, SP, SI, HR, MO, SW, EZ.
Recalling firm
Capso Vision, Inc., Saratoga, CA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1827-2020 | CapsoCAM Plus, UDI: 00867770000209 | 1882 | Class II | Terminated | 6 May 2020 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2019
- 968
- Share of these events in Class II
- 88%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 3
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 26 Feb 2020 | The CapsoCam Plus (SV-3); Device Common Name: System, Imaging, Gastrointestinal, Wireless, Capsule; Catalog Number 2795 | Device | Class II | 1 |
| 3 Jul 2019 | CapsoCam Plus *** System, Imaging, Gastrointestinal, Wireless, Capsule, Electrical Rating 3.1V DC/10mA REF 2795 - Produc | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-1827-2020The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 85378. https://recallregister.com/fda/ii/85378/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
U.S.: CA, TX, FL, VA, PA, SC, HI, DC, NJ, UT, LA. Foreign: FR, AR, IT, UK, BE, TC, AU, CO, GM, GR, SP, SI, HR, MO, SW, EZ.
- What quantity of product is in FDA event 85378?
- The FDA record gives this quantity:
1882
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2019 have the same class?
- 88% of the 968 device events that started in 2019 are in Class II.