Device recall: Beckman Coulter Access Immunoassay Systems, Access Free T4 Thyroxine, REF 33880 (UDI:15099590225834) for use with…

Class II FDA enforcement event 84862. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Beckman Coulter Inc. started this device recall on 30 Apr 2019. The firm address in the FDA record is in Brea, CA. FDA put the recall in Class II. FDA event 84862 has 6 product records. The first FDA enforcement report date is 10 Jun 2020. The record gives the status Terminated, with the termination date 28 Dec 2021.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Distribution: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky,…

Recalling firm

Beckman Coulter Inc., Brea, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

6 product records in the event.

FDA event 84862: product records
Recall numberProductQuantityClassStatusReport date
Z-2237-2020Beckman Coulter Access Immunoassay Systems, Access Free T4 Thyroxine, REF 33880 (UDI:15099590225834) for use with Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.1,322,438 unitsClass IITerminated10 Jun 2020
Z-2238-2020Beckman Coulter Access Immunoassay Systems, Access Free T3 Triiodothyronine, REF A13422 (UDI: 15099590201661) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.418,400 unitsClass IITerminated10 Jun 2020
Z-2239-2020Beckman Coulter ACCESS Immunoassay Systems, Access Total T3 Triiodothyronine , REF 33830 (UDI: 15099590227210) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.238,862Class IITerminated10 Jun 2020
Z-2240-2020Beckman Coulter ACCESS Immunoassay Systems, Access Thyroglobulin, REF 33860 (UDI: 15099590227173) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.95,453 unitsClass IITerminated10 Jun 2020
Z-2241-2020Beckman Coulter ACCESS Immunoassay Systems, Access GI Monitor Cancer Antigen 19-9, REF 387687 ( UDI: 15099590231842) the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.150,887Class IITerminated10 Jun 2020
Z-2242-2020Beckman Coulter ACCESS Immunoassay Systems, Access Thyroglobulin Antibody II, REF A32898 (UDI: 15099590202910) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.123,324 unitsClass IITerminated10 Jun 2020

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2019
968
Share of these events in Class II
88%
Product records in this event
6Median of the firm: 1
Events of the firm on the register
216

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 84862. https://recallregister.com/fda/ii/84862/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Distribution: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky,…
What quantity of product is in FDA event 84862?
The event has 6 product records. The first record gives this quantity: 1,322,438 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2019 have the same class?
88% of the 968 device events that started in 2019 are in Class II.