Device recall: MicroVention Terumo HydroSoft 10, HydroCoil Embolic System, Endovascular Embolization Coil, Helical, Sterile, Rx, REF…
Class II FDA enforcement event 84727. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
The devices may be missing the implant coil.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Microvention, Inc. started this device recall on 22 Nov 2019. The firm address in the FDA record is in Aliso Viejo, CA. FDA put the recall in Class II. FDA event 84727 has 16 product records. The first FDA enforcement report date is 4 Mar 2020. The record gives the status Terminated, with the termination date 26 Jan 2022.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia,…
Recalling firm
Microvention, Inc., Aliso Viejo, CA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
16 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1392-2020 | MicroVention Terumo HydroSoft 10, HydroCoil Embolic System, Endovascular Embolization Coil, Helical, Sterile, Rx, REF numbers 100101H2HS-V; 100102H2HS-V, 100103H2HS-V, 100104H2HS-V, 100151H2HS-V; 100152H2HS-V; 100153H2HS-V; 100154H2HS-V; 100201H2HS-V; 100202H2HS-V; 100203H2HS-V; 100204H2HS-V; 100206H2HS-V,… | 2,762 units | Class II | Terminated | 4 Mar 2020 |
| Z-1393-2020 | MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, HyperSoft 3D, Sterile, Rx, REF numbers 100102HS3D-V, 100103HS3D-V, 100152HS3D-V, 100153HS3D-V, 100154HS3D-V, 100202HS3D-V, 100203HS3D-V, 100204HS3D-V, 100206HS3D-V, 100208HS3D-V, 100254HS3D-V, 100256HS3D-V, 100304HS3D-V,… | 1,725 units | Class II | Terminated | 4 Mar 2020 |
| Z-1394-2020 | MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Helical 10 Regular, Sterile, Rx, REF numbers 100515HC-R-V, 100515HCSR-R-V, 100520HC-R-V, 100520HCSR-R-V, 100620HC-R-V, 100620HCSR-R-V, 100730HC-R-V, 100730HCSR-R-V, 100830HC-R-V, 100930HC-R-V, 100930HCSR-R-V, 101030HC-R-V, and… | 486 units | Class II | Terminated | 4 Mar 2020 |
| Z-1395-2020 | MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Helical 10 Soft, Sterile, Rx, REF numbers 100206HCSR-S-V, 100208HC-S-V, 100208HCSR-S-V, 100306HCSR-S-V, 100308HCSR-S-V, 100310HC-S-V, 100310HCSR-S-V, 100410HC-S-V, 100410HCSR-S-V, Product Usage: The MCS is also intended for… | 780 units | Class II | Terminated | 4 Mar 2020 |
| Z-1396-2020 | MicroVention Terumo Microplex 18, Platinum Coil System, Endovascular Embolization Coil, Helical 18 Regular, Sterile, Rx, REF numbers 180520HC-R-V and 181230HC-R-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an… | 9 units | Class II | Terminated | 4 Mar 2020 |
| Z-1397-2020 | MicroVention Terumo Microplex 18, Platinum Coil System, Endovascular Embolization Coil, Helical 18 Soft, Sterile, Rx, REF numbers 180204HC-S-V, 180208HC-S-V, 180308HC-S-V, 180410HC-S-V, and 180512HC-S-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to… | 169 units | Class II | Terminated | 4 Mar 2020 |
| Z-1398-2020 | MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Cosmos 10, Sterile, Rx, REF numbers 100202CSSR-V, 100254CSSR-V, 100306CSSR-V, 100408CSSR-V, 100412CSSR-V, 100515CSSR-V, 100522CSSR-V, 100618CSSR-V, 100626CSSR-V, 100722CSSR-V, 100731CSSR-V, 100825CSSR-V, 100837CSSR-V,… | 3,442 units | Class II | Terminated | 4 Mar 2020 |
| Z-1399-2020 | MicroVention Terumo Microplex 18, Platinum Embolization Coil, Cosmos 18, Sterile, Rx, REF numbers 180619CS-V, 180723CS-V, 180827CS-V, 180931CS-V, 181036CS-V, 181139CS-V, 181243CS-V, 181347CS-V, 181451CS-V, 181652CS-V, 181859CS-V, 182065CS-V, 182263CS-V, and 182468CS-V; Product Usage: The MCS is also intended for… | 403 units | Class II | Terminated | 4 Mar 2020 |
| Z-1400-2020 | MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Complex 10, Sterile, Rx, REF numbers 100204CC-V, 100307CC-V, 100410CC-V, 100512CC-V, 100615CC-V, 100718CC-V, 100820CC-V, 100924CC-V, and 101026CC-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels… | 409 units | Class II | Terminated | 4 Mar 2020 |
| Z-1401-2020 | MicroVention Terumo Microplex 18, Platinum Coil System, Endovascular Embolization Coil, Complex 18, Sterile, Rx, REF numbers 180512CC-V, 180820CC-V, 181026CC-V, 181434CC-V, and 182050CC-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently… | 19 units | Class II | Terminated | 4 Mar 2020 |
| Z-1402-2020 | MicroVention Terumo HydroFrame 10, HydroCoil Embolic System, Endovascular Embolization Coil, Sterile, Rx, REF numbers 100306HFRM-V, 100405HFRM-V, 100408HFRM-V, 100410HFRM-V, 100412HFRM-V, 100415HFRM-V, 100510HFRM-V, 100515HFRM-V, 100519HFRM-V, 100612HFRM-V, 100619HFRM-V, 100623HFRM-V, 100715HFRM-V, 100728HFRM-V,… | 991 units | Class II | Terminated | 4 Mar 2020 |
| Z-1403-2020 | MicroVention Terumo HydroFrame 18, HydroCoil Embolic System, Endovascular Embolization Coil, Sterile, Rx, REF numbers 180723HFRM-V, 180931HFRM-V, 181644HFRM-V, MV-80619HHFA, MV80827HHFA, MV-80931HHFA, and MV-81036HHFA. Product Usage: The HES is also intended for vascular occlusion of blood vessels within the… | 32 units | Class II | Terminated | 4 Mar 2020 |
| Z-1404-2020 | MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Compass 10, Sterile, Rx, REF numbers 100203CMSR-V, 100408CMSR-V, 100412CMSR-V, 100510CMSR-V, 100516CMSR-V, 100612CMSR-V, 100618CMSR-V, and 100721CMSR-V. Product Usage: The MCS is also intended for vascular occlusion of blood… | 85 units | Class II | Terminated | 4 Mar 2020 |
| Z-1405-2020 | MicroVention Terumo HydroSoft 10, HydroCoil Embolic System, Endovascular Embolization Coil, HydroSoft 3D, Sterile, Rx, REF numbers 100102HFRM-V, 100152HFRM-V, 100153HFRM-V, 100154HFRM-V, 100202HFRM-V, 100206HFRM-V, 100208HFRM-V, 100254HFRM-V, 100256HFRM-V, 100308HFRM-V, MV-00202HHTA, MV-00203HHTA, MV-00204HHTA,… | 847 units | Class II | Terminated | 4 Mar 2020 |
| Z-1406-2020 | MicroVention Terumo Microplex, Platinum Coil System, Endovascular Embolization Coil, VFC, Sterile, Rx, REF numbers MV-00306VFC, MV-00315VFC, MV-00310VFC, MV-00630VFC, VFC010303-V, VFC030606-V, VFC030610-V, VFC030615-V, VFC061020-V, VFC061030-V, VFC101530-V, VFC101540-V, VFC152040-V, and VFC152060-V. Product Usage: .… | 666 units | Class II | Terminated | 4 Mar 2020 |
| Z-1407-2020 | MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, HyperSoft Helical, Sterile, Rx, REF numbers 100101HS-V, 100102HS-V, 100103HS-V, 100104HS-V, 100151HS-V, 100152HS-V, 100153HS-V, 100154HS-V, 100201HS-V, 100202HS-V, 100203HS-V, 100204HS-V, 100206HS-V, 100254HS-V, 100256HS-V,… | 5,777 units | Class II | Terminated | 4 Mar 2020 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2019
- 968
- Share of these events in Class II
- 88%
- Product records in this event
- 16Median of the firm: 1
- Events of the firm on the register
- 16
Other events of the firm
Official channels
- FDA enforcement report Z-1392-2020The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 84727. https://recallregister.com/fda/ii/84727/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia,…
- What quantity of product is in FDA event 84727?
- The event has 16 product records. The first record gives this quantity:
2,762 units
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2019 have the same class?
- 88% of the 968 device events that started in 2019 are in Class II.