Device recall: da Vinci Xi Surgical System Surgical System, IS4000, 00886874114216 - Product Usage: used with an endoscope, surgical…

Class II FDA enforcement event 84335. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (PSC). As a result, you may experience either vision loss or non-recoverable errors 307 or 319, preventing further use of the system.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Intuitive Surgical, Inc. started this device recall on 26 Jul 2019. The firm address in the FDA record is in Sunnyvale, CA. FDA put the recall in Class II. FDA event 84335 has 3 product records. The first FDA enforcement report date is 12 Feb 2020. The record gives the status Terminated, with the termination date 10 Aug 2022.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide distribution in the states of Alabama, Alaska, Arizona, California, Colorado, Connecticut, Florida, Hawaii, Idaho, Illinois, Iowa, Kansas, Kentucky, Louisiana,…

Recalling firm

Intuitive Surgical, Inc., Sunnyvale, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

3 product records in the event.

FDA event 84335: product records
Recall numberProductQuantityClassStatusReport date
Z-0925-2020da Vinci Xi Surgical System Surgical System, IS4000, 00886874114216 - Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.123 surgical systemsClass IITerminated12 Feb 2020
Z-0926-2020da Vinci X Surgical System Surgical System, IS4200, 00886874114216 Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.52 surgical systemsClass IITerminated12 Feb 2020
Z-0927-2020da Vinci SP Surgical System, SP1098, 00886874114605 Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.4 Surgical SystemsClass IITerminated12 Feb 2020

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2019
968
Share of these events in Class II
88%
Product records in this event
3Median of the firm: 1
Events of the firm on the register
105

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 84335. https://recallregister.com/fda/ii/84335/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide distribution in the states of Alabama, Alaska, Arizona, California, Colorado, Connecticut, Florida, Hawaii, Idaho, Illinois, Iowa, Kansas, Kentucky, Louisiana,…
What quantity of product is in FDA event 84335?
The event has 3 product records. The first record gives this quantity: 123 surgical systems
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2019 have the same class?
88% of the 968 device events that started in 2019 are in Class II.