Device recall: UniCel DxH Slidemaker Stainer (DxH SMS) and UniCel DxH Slidemaker Stainer II (DxH SMS II) - Product Usage: The…

Class II FDA enforcement event 84186. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

When the UniCel DxH Slidemaker Stainer and UniCel DxH Slidemaker Stainer II dispense probe is slightly bent, there is a potential for cell carryover from the previous sample, and erroneous results could be reported from the slide review. Erroneous results may be reported from the laboratory which may trigger unnecessary or inappropriate patient treatment.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Beckman Coulter, Inc. started this device recall on 7 Oct 2019. The firm address in the FDA record is in Miami, FL. FDA put the recall in Class II. FDA event 84186 has 1 product record. The first FDA enforcement report date is 4 Dec 2019. The record gives the status Terminated, with the termination date 27 Apr 2021.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution - US Nationwide distribution and countries of Australia, Bahrain, Belarus, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Egypt, Ethiopia, France, Germany, Greece,…

Recalling firm

Beckman Coulter, Inc., Miami, FL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 84186: product records
Recall numberProductQuantityClassStatusReport date
Z-0516-2020UniCel DxH Slidemaker Stainer (DxH SMS) and UniCel DxH Slidemaker Stainer II (DxH SMS II) - Product Usage: The intended use is as follows: The DxH Slidemaker Stainer is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and…1,056 (418 US, and 638 OUS)Class IITerminated4 Dec 2019

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2019
968
Share of these events in Class II
88%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
216

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 84186. https://recallregister.com/fda/ii/84186/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution - US Nationwide distribution and countries of Australia, Bahrain, Belarus, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Egypt, Ethiopia, France, Germany, Greece,…
What quantity of product is in FDA event 84186?
The FDA record gives this quantity: 1,056 (418 US, and 638 OUS)
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2019 have the same class?
88% of the 968 device events that started in 2019 are in Class II.