Device recall: CS5/5+ Fastpacks, 225 mL, 150¿ Res Autotransfusion Device - Product Usage: its related accessory components are…
Class II FDA enforcement event 84183. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
Potential to develop leaks at the inner core of Cell Saver 5/5+ and Cell Saver Elite/Elite +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Haemonetics Corporation started this device recall on 21 Oct 2019. The firm address in the FDA record is in Braintree, MA. FDA put the recall in Class II. FDA event 84183 has 22 product records. The first FDA enforcement report date is 5 Feb 2020. The record gives the status Terminated, with the termination date 6 Nov 2020.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM,…
Recalling firm
Haemonetics Corporation, Braintree, MA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
22 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0885-2020 | CS5/5+ Fastpacks, 225 mL, 150¿ Res Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. | 44821 | Class II | Terminated | 5 Feb 2020 |
| Z-0886-2020 | CS5/5+ Bowl Kit-125 mL Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. | 333443 | Class II | Terminated | 5 Feb 2020 |
| Z-0887-2020 | High Speed Cell Saver5/5+ Bowl Set Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. | 3563 | Class II | Terminated | 5 Feb 2020 |
| Z-0888-2020 | CS5/5+ Fastpacks, 125 mL, 150¿ Res Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. | 14095 | Class II | Terminated | 5 Feb 2020 |
| Z-0889-2020 | CS5/5+ Fastpacks, 125 mL, 20¿ Res Autotransfusion Device- Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. | 1340 | Class II | Terminated | 5 Feb 2020 |
| Z-0890-2020 | CS5/5+ Fastpacks, 225 mL, 20¿ Res Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. | 8641 | Class II | Terminated | 5 Feb 2020 |
| Z-0891-2020 | Cell Saver5/5+ Bowl Kit (225 ml) Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. | 690043 | Class II | Terminated | 5 Feb 2020 |
| Z-0892-2020 | CS Elite Processing Set (225 ml) Autotransfusion Device - Product Usage:its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product… | 257345 | Class II | Terminated | 5 Feb 2020 |
| Z-0893-2020 | CS Elite Processing Set (125 ml) Autotransfusion Device - Product Usage:its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product… | 176742 | Class II | Terminated | 5 Feb 2020 |
| Z-0894-2020 | CS Elite Fastpack, 225 mL, 20¿ Res Autotransfusion Device - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell… | 1092 | Class II | Terminated | 5 Feb 2020 |
| Z-0895-2020 | CS Elite Fastpack, 225 mL, 150 Res Autotransfusion Device - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell… | 42739 | Class II | Terminated | 5 Feb 2020 |
| Z-0896-2020 | CS Elite Fastpack, 125 mL, 150 Res Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. | 14698 | Class II | Terminated | 5 Feb 2020 |
| Z-0897-2020 | JAPAN LOW VOL(125ML) CS5 SET,LN 261J - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. The recovered blood is processed by a centrifugation and washing procedure before pumping the red… | 38699 | Class II | Terminated | 5 Feb 2020 |
| Z-0898-2020 | JAPAN BASIC HI SPEED CS5 SET,LN 263J - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. The recovered blood is processed by a centrifugation and washing procedure before pumping the red… | 28462 | Class II | Terminated | 5 Feb 2020 |
| Z-0899-2020 | JAPAN DISP SET, 261J, 205J, 208 - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. The recovered blood is processed by a centrifugation and washing procedure before pumping the red cell… | 25529 | Class II | Terminated | 5 Feb 2020 |
| Z-0900-2020 | JAPAN DISP SET, 261J, 205J, HAR - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. The recovered blood is processed by a centrifugation and washing procedure before pumping the red cell… | 739 | Class II | Terminated | 5 Feb 2020 |
| Z-0901-2020 | Japan CS Elite FASTPACK DISP. SET - 125ML - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a… | 14278 | Class II | Terminated | 5 Feb 2020 |
| Z-0902-2020 | JAPAN CS ELITE FASTPACK, 225ML - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for… | 8114 | Class II | Terminated | 5 Feb 2020 |
| Z-0903-2020 | JAPAN CELL SAVER ELITE SET - 125ML - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for… | 21113 | Class II | Terminated | 5 Feb 2020 |
| Z-0904-2020 | JAPAN CELL SAVER ELITE SET - 225ML - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for… | 12370 | Class II | Terminated | 5 Feb 2020 |
| Z-0905-2020 | Japan DISP SET, 263J, 205J, 208 - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma | 14271 | Class II | Terminated | 5 Feb 2020 |
| Z-0906-2020 | Japan DISP SET, 263J, 205J, HAR - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma | 1584 | Class II | Terminated | 5 Feb 2020 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2019
- 968
- Share of these events in Class II
- 88%
- Product records in this event
- 22Median of the firm: 1
- Events of the firm on the register
- 8
Other events of the firm
Official channels
- FDA enforcement report Z-0885-2020The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 84183. https://recallregister.com/fda/ii/84183/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM,…
- What quantity of product is in FDA event 84183?
- The event has 22 product records. The first record gives this quantity:
44821
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2019 have the same class?
- 88% of the 968 device events that started in 2019 are in Class II.