Device recall: Infusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as…

Class II FDA enforcement event 83549. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Potential for fluid leakage out of air vents on inline filters on Plum and Sapphire micropore infusion sets.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

ICU Medical Inc started this device recall on 29 Jul 2019. The firm address in the FDA record is in Lake Forest, IL. FDA put the recall in Class II. FDA event 83549 has 22 product records. The first FDA enforcement report date is 23 Oct 2019. The record gives the status Terminated, with the termination date 12 Jan 2024.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD,…

Recalling firm

ICU Medical Inc, Lake Forest, IL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

22 product records in the event.

FDA event 83549: product records
Recall numberProductQuantityClassStatusReport date
Z-0126-2020Infusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as High-Pressure Filter PRIMARY I.V. PLUMSET Convetible Pin, 104 Inch with Orange Polyethylene-Lined*/Light Resistant Tubing, Prepierced Injection Site and OPTION-LOK. 1 unit per pouch; 48 pouches per case.170,735 unitsClass IITerminated23 Oct 2019
Z-0127-2020Infusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as Primary PLUM Set 0.2 Micron Filter, Prepierced Y-Site, Polyethylene Lined Light Resistant Tubing, Distal Microbore Tubing, Secure Lock, 104 Inch. 1 unit per pouch; 50 pouches per case78,877 unitsClass IITerminated23 Oct 2019
Z-0128-2020Infusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as Primary PLUM Set Yellow Striped Tubing, Distal Microbore Tubing, 0.2 Micron Filter, Secure Lock, 107 Inch. 1 unit per pouch; 48 pouches per case1,536 unitsClass IITerminated23 Oct 2019
Z-0129-2020Infusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Primary 1.2 Micron Filter Set Microbore, 122 Inch. 1 unit per pouch; 50 pouches per case73,710 unitsClass IITerminated23 Oct 2019
Z-0130-2020Infusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Primary 0.2 Micron Filter Set, Microbore, CLAVE Y-Site, 118 Inch. 1 unit per pouch; 50 pouches per case45,900 unitsClass IITerminated23 Oct 2019
Z-0131-2020Infusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Epidural Set Yellow-Striped Microbore, 0.2 Micron Filter 119 Inch. 1 unit per pouch; 50 pouches per case155,150 unitsClass IITerminated23 Oct 2019
Z-0132-2020Infusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Epidural Set Yellow-Striped Microbore, 0.2 Micron Filter, PAV 123 inch. 1 unit per pouch; 50 pouches per case26,700 unitsClass IITerminated23 Oct 2019
Z-0133-2020Infusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Primary 0.2 Micron Filter Set Non-Vented, Microbore, PAV, CLAVE Y-Site, 122 Inch. 1 unit per pouch; 50 pouches per case51,250 unitsClass IITerminated23 Oct 2019
Z-0134-2020Infusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Primary 1.2 Micron Filter Set Non-Vented, Microbore, PAV, 123 Inch. 1 unit per pouch; 50 pouches per case19,350 unitsClass IITerminated23 Oct 2019
Z-0135-2020Infusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Epidural Set with NRFit Connector Yellow-Striped Microbore, 0.2 Micron Filter 119 Inch. 1 unit per pouch; 50 pouches per case4,600 unitsClass IITerminated23 Oct 2019
Z-0136-2020Infusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as PRIMARY PLUMSET 15 Micron Filter in Sight Chamber, CLAVE Port, 0.2 Micron Filter, CLAVE Y-Site, PE Lined Light Resistant Tubing, Distal Microbore Tubing 272 cm / 15 mL. 1 unit per pouch; 50 pouches per case.…14,850 unitsClass IITerminated23 Oct 2019
Z-0137-2020Infusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as Primary PLUM Set 15 Micron Filter in Sight Chamber, CLAVE Secondary Port, 0.2 Micron Filter, CLAVE Y-Site, Polyethylene Lined Light Resistant Tubing, Distal Microbore Tubing, Secure Lock 272 cm / 15 mL. 1 unit…280,826 unitsClass IITerminated23 Oct 2019
Z-0138-2020Infusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as Primary PLUM Set 15 Micron Filter in Sight Chamber, CLAVE Secondary Port, 0.2 Micron Filter, CLAVE Y-Site, Polyethylene Lined Light Resistant Tubing, Distal Microbore Tubing, Secure Lock 272 cm / 15 mL. 1 unit…6,400 unitsClass IITerminated23 Oct 2019
Z-0139-2020Infusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as Primary PLUM Set, 15 Micron Filter in Sight Chamber, CLAVE Secondary Port, 0.2 Micron Filter, Polyethylene Lined Light Resistant Tubing, Distal Microbore Tubing, Secure Lock 269 cm / 14 mL. 1 unit per pouch; 50…521 unitsClass IITerminated23 Oct 2019
Z-0140-2020Infusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Primary 1.2 Micron Filter Set Non-Vented, Microbore 311 cm / 8.4 mL. 1 unit per pouch; 1 unit per blister; 40 blisters per case . Not distributed in the U.S.2,327 unitsClass IITerminated23 Oct 2019
Z-0141-2020Infusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Primary 1.2 Micron Filter Set Non-Vented, Microbore 311 cm / 8.4 mL. 1 unit per pouch; 1 unit per blister; 40 blisters per case . Not distributed in the U.S.32,416 unitsClass IITerminated23 Oct 2019
Z-0142-2020Infusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Primary 0.2 Micron Filter Set Non-Vented, Microbore, 293 cm / 8.1 mL. 1 unit per pouch; 1 unit per blister; 40 blisters per case. Not distributed in the U.S.5,443 unitsClass IITerminated23 Oct 2019
Z-0143-2020Infusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Epidural Set Yellow-Striped Microbore, 0.2 Micron Filter 301 cm / 8.3 mL. 1 unit per blister; 40 blisters per case . Not distributed in the U.S.400 unitsClass IITerminated23 Oct 2019
Z-0144-2020Infusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Epidural Set Yellow-Striped Microbore, 0.2 Micron Filter 301 cm / 8.3 mL. 1 unit per blister; 40 blisters per case . Not distributed in the U.S.6,516 unitsClass IITerminated23 Oct 2019
Z-0145-2020Infusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as PLUM Epidural Set with NRFit Connector Yellow-Striped Tubing, Distal Microbore Tubing, 0.2 Micron Filter, 107 Inch. 1 unit per pouch; 48 pouches per case7,920 unitsClass IITerminated23 Oct 2019
Z-0146-2020Infusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Epidural Set with NRFit Connector Yellow-Striped Microbore, 0.2 Micron Filter 301 cm / 8.3 mL. 1 unit per blister; 40 blisters per case. Not distributed in the U.S.120 unitsClass IITerminated23 Oct 2019
Z-0147-2020Infusion Pump I.V. Administration, specifically Sapphire Microbore Infusion Sets with Inline Filters labeled as SAPPHIRE Epidural Set with NRFit Connector Yellow-Striped Microbore, 0.2 Micron Filter, PAV, 123 Inch. 1 unit per pouch; 50 pouches per case50 unitsClass IITerminated23 Oct 2019

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2019
968
Share of these events in Class II
88%
Product records in this event
22Median of the firm: 2
Events of the firm on the register
75

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 83549. https://recallregister.com/fda/ii/83549/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD,…
What quantity of product is in FDA event 83549?
The event has 22 product records. The first record gives this quantity: 170,735 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2019 have the same class?
88% of the 968 device events that started in 2019 are in Class II.