Recall Register

Device recall: Plum 360 Infusion System With MedNet/ Smart Card Plug And Play Module, Model # 30010. Infusion pump.

Class II FDA enforcement event 83480. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

A software malfunction of the infuser when used with a MedNet Custom Drug Library (CDL). When a new CDL is downloaded and installed on the infuser the infuser can become inoperable and show a blank screen on the User Interface.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

ICU Medical Inc started this device recall on 1 Aug 2018. The firm address in the FDA record is in Lake Forest, IL. FDA put the recall in Class II. FDA event 83480 has 1 product record. The first FDA enforcement report date is 30 Oct 2019. The record gives the status Terminated, with the termination date 27 Apr 2021.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US nationwide distribution, and Canada

Recalling firm

ICU Medical Inc, Lake Forest, IL.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 83480: product records
Recall numberProductQuantityClassStatusReport date
Z-0171-2020Plum 360 Infusion System With MedNet/ Smart Card Plug And Play Module, Model # 30010. Infusion pump.47,152 devicesClass IITerminated30 Oct 2019

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2018
1,074
Share of these events in Class II
92%
Product records in this event
1Median of the firm: 2
Events of the firm on the register
75

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 83480. https://recallregister.com/fda/ii/83480/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US nationwide distribution, and Canada
What quantity of product is in FDA event 83480?
The FDA record gives this quantity: 47,152 devices
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2018 have the same class?
92% of the 1,074 device events that started in 2018 are in Class II.