Device recall: Radrex-i X-Ray System (DRAD-3000E) TFP-4336W (Wireless FPD) Software version: v5.00ER005 Product Usage: The RADREX-I…
Class II FDA enforcement event 83193. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
software malfunction; It was found when a user performs radiography using the wireless flat panel detector (FPD), a message window displays on the monitor stating image transmission was not completed and there was no image available. It also showed the "OK" button to reacquire image data from the FPD, and the "Cancel" button to cancel the acquisition . When the user selects the "OK" button, the…
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Canon Medical System, USA, INC. started this device recall on 16 Oct 2018. The firm address in the FDA record is in Tustin, CA. FDA put the recall in Class II. FDA event 83193 has 1 product record. The first FDA enforcement report date is 17 Jul 2019. The record gives the status Terminated, with the termination date 20 Jul 2020.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US Nationwide Distribution - PA, UT, FL, NY, AR, LA, WV, NJ, OH, MI, WI, TN, TX, VT, GA, NC, IL
Recalling firm
Canon Medical System, USA, INC., Tustin, CA.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1958-2019 | Radrex-i X-Ray System (DRAD-3000E) TFP-4336W (Wireless FPD) Software version: v5.00ER005 Product Usage: The RADREX-I is a general purpose x-ray system that employs Solid State Imager(s), SSXI, which converts x-rays directly into electrical signals which can, after appropriate processing be displayed on LCD monitors… | 46 | Class II | Terminated | 17 Jul 2019 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2018
- 1,074
- Share of these events in Class II
- 92%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 27
Other events of the firm
Official channels
- FDA enforcement report Z-1958-2019The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 83193. https://recallregister.com/fda/ii/83193/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US Nationwide Distribution - PA, UT, FL, NY, AR, LA, WV, NJ, OH, MI, WI, TN, TX, VT, GA, NC, IL
- What quantity of product is in FDA event 83193?
- The FDA record gives this quantity:
46
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2018 have the same class?
- 92% of the 1,074 device events that started in 2018 are in Class II.