Recall Register

Device recall: BRANSIST safire Product Usage: DIGITAL ANGIOGRAPHY SYSTEM This system is intended to be used for radiologic…

Class II FDA enforcement event 81981. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Normal operation of the device is to power up the device in the morning, register the first patient, and then perform a fluoroscopy. If, however, the user powers up the device in the morning and makes an error by starting a fluoroscopy while the first patient is still being registered, the device application will abnormally terminate and require service intervention before it can be used again.…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Shimadzu Medical Systems Usa Com started this device recall on 14 Jan 2019. The firm address in the FDA record is in Torrance, CA. FDA put the recall in Class II. FDA event 81981 has 1 product record. The first FDA enforcement report date is 6 Mar 2019. The record gives the status Terminated, with the termination date 7 Jun 2024.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide Distribution in the states of IN, OH, IL, SC

Recalling firm

Shimadzu Medical Systems Usa Com, Torrance, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 81981: product records
Recall numberProductQuantityClassStatusReport date
Z-0959-2019BRANSIST safire Product Usage: DIGITAL ANGIOGRAPHY SYSTEM This system is intended to be used for radiologic visualization of the heart, blood vessels or lymphatic system during or after injection of a contrast medium. It is to be used in the diagnosis of the circulatory vascular system while being combined with a…8 unitsClass IITerminated6 Mar 2019

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2019
968
Share of these events in Class II
88%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
5

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 81981. https://recallregister.com/fda/ii/81981/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide Distribution in the states of IN, OH, IL, SC
What quantity of product is in FDA event 81981?
The FDA record gives this quantity: 8 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2019 have the same class?
88% of the 968 device events that started in 2019 are in Class II.