Device recall: AVE 2 Birthing Bed
Class II FDA enforcement event 81927. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
LINET has recently become aware of a potential safety issue for the AVE 2 Birthing Bed caused by the power cable breaking at the bed connection or the orange insulation of the power cable may fail causing a possible spark or possible exposure to a live conductor.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Linet Spol. S.r.o. started this device recall on 7 Jan 2019. The firm address in the FDA record is in Slany. FDA put the recall in Class II. FDA event 81927 has 1 product record. The first FDA enforcement report date is 20 Mar 2019. The record gives the status Terminated, with the termination date 31 May 2023.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
AL, OK, NJ, OR, WA, CA, MN, LA, AZ, IL, MO, CO, OH, KY, FL, KS, UT, TN, LA, DE, CT, TX
Recalling firm
Linet Spol. S.r.o., Slany, N/A, Czech Republic.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
1 product record in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-0991-2019 | AVE 2 Birthing Bed | 253 devices | Class II | Terminated | 20 Mar 2019 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2019
- 968
- Share of these events in Class II
- 88%
- Product records in this event
- 1Median of the firm: 1
- Events of the firm on the register
- 3
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 18 Jan 2023 | Sprint 200 wheeled stretcher, Item numbers 1ES200-7, 1ES210-1, and 1ES211-1. | Device | Class II | 1 |
| 13 Nov 2019 | Eleganza 5 (AC powered adjustable bed), Item Numbers: a) 1GE512055-40, b) 1GE512055-45, c) 1GE512055-85, and d) 1GE51205 | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-0991-2019The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 81927. https://recallregister.com/fda/ii/81927/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
AL, OK, NJ, OR, WA, CA, MN, LA, AZ, IL, MO, CO, OH, KY, FL, KS, UT, TN, LA, DE, CT, TX
- What quantity of product is in FDA event 81927?
- The FDA record gives this quantity:
253 devices
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2019 have the same class?
- 88% of the 968 device events that started in 2019 are in Class II.