Recall Register

Device recall: Parietex Composite Parastomal Mesh (Product Number PCOPM15 / PCOPM20) Product Usage: The PCOPM15 and PCOPM20 reorder…

Class II FDA enforcement event 81417. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The firm is issuing this voluntary removal following receipt of reports of parastomal mesh failure identified several years following parastomal hernia repair using the modified Sugarbaker repair technique. In these reports, product failure led to hernia recurrence requiring additional surgical treatment. Symptoms of hernia recurrence may include discomfort, localized pain-free or painful…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Covidien LLC started this device recall on 5 Oct 2018. The firm address in the FDA record is in North Haven, CT. FDA put the recall in Class II. FDA event 81417 has 1 product record. The first FDA enforcement report date is 21 Nov 2018. The record gives the status Terminated, with the termination date 7 May 2020.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Worldwide distribution--US Nationwide

Recalling firm

Covidien LLC, North Haven, CT.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 81417: product records
Recall numberProductQuantityClassStatusReport date
Z-0461-2019Parietex Composite Parastomal Mesh (Product Number PCOPM15 / PCOPM20) Product Usage: The PCOPM15 and PCOPM20 reorder codes are circular in shape, made from a three-dimensional monofilament polyester fabric with a two-dimensional monofilament polyester central band. One side of the design is completely covered with…7333Class IITerminated21 Nov 2018

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2018
1,074
Share of these events in Class II
92%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
77

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 81417. https://recallregister.com/fda/ii/81417/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Worldwide distribution--US Nationwide
What quantity of product is in FDA event 81417?
The FDA record gives this quantity: 7333
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2018 have the same class?
92% of the 1,074 device events that started in 2018 are in Class II.