Recall Register

Device recall: 26-Lead Transfer Set for use with APEX Compounding System, Product Catalog Number 2112550

Class II FDA enforcement event 80231. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

There is a potential for valves on the 26-Lead Transfer Set to be incorrectly oriented. When a transfer set with an incorrectly oriented valve is installed on the system, there is potential for unintended admixture.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

B. Braun Medical, Inc. started this device recall on 4 Jun 2018. The firm address in the FDA record is in Allentown, PA. FDA put the recall in Class II. FDA event 80231 has 1 product record. The first FDA enforcement report date is 11 Jul 2018. The record gives the status Terminated, with the termination date 13 Jan 2020.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

The products were distributed to the following US states: AL, CA, CO, CT, FL, GA, IL, LA, MA, MD, MI, MN, MO, NC, NJ, NV, OH, PA, TX, and WA.

Recalling firm

B. Braun Medical, Inc., Allentown, PA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 80231: product records
Recall numberProductQuantityClassStatusReport date
Z-2343-201826-Lead Transfer Set for use with APEX Compounding System, Product Catalog Number 211255012320Class IITerminated11 Jul 2018

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2018
1,074
Share of these events in Class II
92%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
90

Other events of the firm

Other FDA enforcement events of B. Braun Medical, Inc.
InitiatedProduct (first record)TypeClassRecords
19 Aug 20260.9% Sodium Chloride Injection, USP, 100 mL Partial Fill in 150 mL PAB Container, Sterile, Rx only, B. Braun Medical IncDrugClass I1
12 Aug 2026Lactated Ringer's Injection USP, 1000 mL EXCEL container, Rx only, L7500, B. Braun Medical, Inc., Bethlehem, PA 18018-35DrugClass I1
4 Aug 2026Brand Name: Extension Set Product Name: DISCOFIX FOUR WAY STOPCOCK Model/Catalog Number: 456025DeviceClass II1
6 May 2026Spinal Needle procedure kits: Material Description (Material Number) 24GA. PENCAN SPINAL BUP. TRAY W/DURAPREP (333867); DeviceClass I3
28 Apr 2026Lactated Ringer's Injection USP, 1000 mL container, Rx only, B. Braun Medical, Inc., Bethlehem, PA 18018-3524 USA, NDC 0DrugClass I1
2 Apr 2026Sterile Water for Injection USP, 3000 mL bags, Rx only, Pharmacy Bulk Package, B. Braun Medical Inc., Bethlehem, PA 1801DrugClass II4
19 Mar 2026B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG DR. Model Number: SL-2010M2096A.DeviceClass I6
2 Feb 2026Brand Name: B. Braun Product Name: BBraun Medical 21G X4.4CM Winged INF Model/Catalog Number: 7B3050 Software Version: NDeviceClass II3
29 Jan 2026Brand Name: B. BRAUN MEDICAL INC Product Name: CESK NORTHSIDE ANESTHESIA TRAY Model/Catalog Number: 555076 Software VersDeviceClass II8
29 Oct 2025ADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space LarDeviceClass II116

B. Braun Medical, Inc.: all 90 events

Official channels

Cite this page

Recall Register. FDA enforcement event 80231. https://recallregister.com/fda/ii/80231/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: The products were distributed to the following US states: AL, CA, CO, CT, FL, GA, IL, LA, MA, MD, MI, MN, MO, NC, NJ, NV, OH, PA, TX, and WA.
What quantity of product is in FDA event 80231?
The FDA record gives this quantity: 12320
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2018 have the same class?
92% of the 1,074 device events that started in 2018 are in Class II.