Recall Register

Device recall: Current, Sterile EO, Model #/ Part #: 1207-30/50020236-001, 60010739-207, 60010739-407; 1207-36/50020237-001…

Class II FDA enforcement event 80212. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against unauthorized device access. For older generation devices not capable of accepting the firmware, the firm is advising customers of the option to permanently disable…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

St Jude Medical Inc. started this device recall on 16 Apr 2018. The firm address in the FDA record is in Sylmar, CA. FDA put the recall in Class II. FDA event 80212 has 12 product records. The first FDA enforcement report date is 4 Jul 2018. The record gives the status Terminated, with the termination date 8 Mar 2024.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

U.S. Nationwide distribution

Recalling firm

St Jude Medical Inc., Sylmar, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

12 product records in the event.

FDA event 80212: product records
Recall numberProductQuantityClassStatusReport date
Z-2318-2018Current, Sterile EO, Model #/ Part #: 1207-30/50020236-001, 60010739-207, 60010739-407; 1207-36/50020237-001, 60010742-207, 60010742-307, 60010742-407, 60010742-707; 2207-30/ 50020238-001, 60010738-207; 2207-36/ 50020239-001, 60010741-207, 60010741-307, 60010741-407, 60010741-707; CD1211-36/100006983, 100012542,…71651Class IITerminated4 Jul 2018
Z-2319-2018Ellipse, Sterile EO, Model #/ Part #: CD1275-36/100116115; CD1275-36Q/100116116; CD1277-36/100048388, 100048391, 100060126, 100116208, 10011636; CD1277-36Q/100048374, 100048394, 100060198, 100098497, 100116268, 100116345; CD1293-36Q/100057838; CD1309-36/100057717; CD1309-36Q/100057718; CD1311-36/100057802, 100060153;…139351Class IITerminated4 Jul 2018
Z-2320-2018Fortify, Sterile EO, Model #/ Part #: CD1231-40/100029238,100030095, 100041981; CD1231-40Q/100029229, 100030087, 100041982; CD1233-40/100029230, 100029250, 100029259,100031058, 100059778, 100070017; CD1233-40Q/100029260, 100029271, 100029277, 100031104, 100070089; CD1235-40/100029272, 100029282, 100029311, 100031039,…81338Class IITerminated4 Jul 2018
Z-2321-2018Fortify Assura, Sterile EO, Model #/ Part #: CD1257-40/100042322, 100059623; CD1257-40Q/100042341, 100059535; CD1259-40/100041914, 100042338, 100059536, 100059646, 100081567; CD1259-40Q/100041931, 100042339, 100059537, 100059620, 100059626, 100078685; CD1263-40/100042948; CD1263-40Q/100042949; CD1357-40C/100080365,…131093Class IITerminated4 Jul 2018
Z-2322-2018Promote, Sterile EO, Model #/ Part #: 3207-30/60010740-207, 60010740-407; 3207-36/60010743-207, 60010743-407; 3213-36/50022126-001, 60013879-307, 60013879-407, 60013879-707; CD3211-36/100006984,100012548, 100012550, 100016680, 100031621, 100035618, 100035619, 100042487, 100047194; CD3211-36Q/100004091,…37485Class IITerminated4 Jul 2018
Z-2323-2018Promote Quadra, Sterile EO, Model # / Part #: CD3221-36/100025054, 60019195; CD3223-36P/100013763; CD3239-40/100029115, 100029254, 100037181; CD3239-40Q/100029116, 100029244, 10003711510425Class IITerminated4 Jul 2018
Z-2324-2018Quadra Assura, Sterile EO, Model #/ Part #: CD3265-40/100042335, 100059846; CD3265-40Q/100042336, 100059847; CD3267-40/100042956, 100043569, 100059961, 100078810; CD3267-40Q/100043132, 100043539, 100059917, 100059956, 100078904; CD3365-40C/100080435, 100080436; CD3365-40Q/100080426, 100080427; CD3367-40/100078813,…72673Class IITerminated4 Jul 2018
Z-2325-2018Quadra Assura MP, Sterile EO, Model #/ Part #: CD3269-40/100043177; CD3269-40Q/100043178; CD3271-40/100043130; CD3271-40Q/100043191; CD3369-40C/100080339, 100080478; CD3369-40Q/100080439, 100080476, 100105528; CD3371-40/100078814, 100079232, 100079513; CD3371-40C/100078639, 100079231, 100079478, 100105493, 100105494;…68117Class IITerminated4 Jul 2018
Z-2326-2018Unify, Sterile EO, Model #/ Part #: CD3231-40/100029240, 100030104, 100041904; CD3231-40Q/100029281, 100030124, 100041905; CD3235-40/100029256, 100029293, 100029307, 100031111, 100037047, 100048897, 100070009; CD3235-40Q/100029257, 100029294, 100029308, 100031059, 100037048, 100070010; CD3281-40Q/100056770;…40738Class IITerminated4 Jul 2018
Z-2327-2018Unify Assura, Sterile EO, Model #/ Part #: CD3257-40/100042333, 100059944; CD3257-40Q/100042334, 100059936; CD3261-40/100048460, 100048546, 100059897, 100059954, 100059996, 100078828; CD3261-40Q/100048568, 100048581, 100059898, 100059955, 100059997, 100078853; CD3357-40C/100080408, 100080424, 100083511;…65048Class IITerminated4 Jul 2018
Z-2328-2018Unify Quadra, Sterile EO, Model #/ Part #: CD3249-40/100036404, 100038435, CD3249-40Q/100036407, 100038436, CD3251-40/100036405, 100038095, 100038226, 100038437, 100038451, 100072563, CD3251-40Q/100036408, 100038214, 100038230, 100038407, 100038442, 100072564, CD3255-40/100038224, CD3255-40Q/10003651422132Class IITerminated4 Jul 2018
Z-2329-2018Merlin PCS programmer (Model # 3650), Software (Model # 3330)71651Class IITerminated4 Jul 2018

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2018
1,074
Share of these events in Class II
92%
Product records in this event
12Median of the firm: 2
Events of the firm on the register
14

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 80212. https://recallregister.com/fda/ii/80212/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: U.S. Nationwide distribution
What quantity of product is in FDA event 80212?
The event has 12 product records. The first record gives this quantity: 71651
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2018 have the same class?
92% of the 1,074 device events that started in 2018 are in Class II.