Device recall: Dimension Gentamicin Flex reagent cartridge, DM GENT, gentamicin enzyme immunoassay, Lot numbers affected are EB8115…

Class II FDA enforcement event 80009. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

The affected lots may exhibit inaccuracy for Quality Control and patient samples at the low end of the Analytical Measurement Range. The bias is due to the calibration curve not showing separation between the Level 1 and Level 2 calibrators. The negative bias observed for patient, QC, and calibrator samples, when using any of the listed lots, has the potential to impact the interpretation of…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Siemens Healthcare Diagnostics, Inc. started this device recall on 19 Apr 2018. The firm address in the FDA record is in Newark, DE. FDA put the recall in Class II. FDA event 80009 has 2 product records. The first FDA enforcement report date is 20 Jun 2018. The record gives the status Terminated, with the termination date 24 Apr 2020.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

Nationwide distribution. Foreign distribution to Argentina, Australia, Brazil, Bahamas, Canada, Germany, India, Japan, Myanmar, New Zealand, Saudi Arabia, Taiwan, and South Africa.

Recalling firm

Siemens Healthcare Diagnostics, Inc., Newark, DE.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 80009: product records
Recall numberProductQuantityClassStatusReport date
Z-2181-2018Dimension Gentamicin Flex reagent cartridge, DM GENT, gentamicin enzyme immunoassay, Lot numbers affected are EB8115, BB8159, and FB8275.7543Class IITerminated20 Jun 2018
Z-2182-2018Dimension Vista Gentamicin Flex reagent cartridge, DV GENT, gentamicin enzyme immunoassay. Affected lots are 17135BC, 17150BF, 17159BD, 17178AC, 17269BD, and 17312BB.20373Class IITerminated20 Jun 2018

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2018
1,074
Share of these events in Class II
92%
Product records in this event
2Median of the firm: 1
Events of the firm on the register
333

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 80009. https://recallregister.com/fda/ii/80009/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: Nationwide distribution. Foreign distribution to Argentina, Australia, Brazil, Bahamas, Canada, Germany, India, Japan, Myanmar, New Zealand, Saudi Arabia, Taiwan, and South Africa.
What quantity of product is in FDA event 80009?
The event has 2 product records. The first record gives this quantity: 7543
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2018 have the same class?
92% of the 1,074 device events that started in 2018 are in Class II.