Device recall: Trident Endoscopic Ultrasonic Biopsy Needle (FNB)
Class II FDA enforcement event 79911. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
The combination of the firmness of the mass, the potential endoscope and elevator position combined with the force required to penetrate the mass may contribute to the breakage of the needle in the human body, which could cause injury.
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Micro-Tech (Nanjing) Co., Ltd. started this device recall on 12 Mar 2018. The firm address in the FDA record is in Nanjing. FDA put the recall in Class II. FDA event 79911 has 2 product records. The first FDA enforcement report date is 23 May 2018. The record gives the status Terminated, with the termination date 30 Aug 2018.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US distribution to one distributor in MI.
Recalling firm
Micro-Tech (Nanjing) Co., Ltd., Nanjing, China.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
2 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1877-2018 | Trident Endoscopic Ultrasonic Biopsy Needle (FNB) | 40 total | Class II | Terminated | 23 May 2018 |
| Z-1878-2018 | Trident Endoscopic Ultrasonic Aspiration Needle (FNA) | 40 total | Class II | Terminated | 23 May 2018 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2018
- 1,074
- Share of these events in Class II
- 92%
- Product records in this event
- 2Median of the firm: 2
- Events of the firm on the register
- 2
Other events of the firm
| Initiated | Product (first record) | Type | Class | Records |
|---|---|---|---|---|
| 20 Sep 2018 | Single-Use Biopsy Forceps, JUMBO ALLIGATOR,3.2mm, REF Catalog No. NBF13-01130230, UPN BF40180 (Pouch level) - BF40181(Ca Event 81176 | Device | Class II | 1 |
Official channels
- FDA enforcement report Z-1877-2018The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 79911. https://recallregister.com/fda/ii/79911/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US distribution to one distributor in MI.
- What quantity of product is in FDA event 79911?
- The event has 2 product records. The first record gives this quantity:
40 total
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2018 have the same class?