Device recall: Baxter Sigma Spectrum Infusion Pump with master Drug Library (Version 6) 35700BAX Product Usage: The Baxter Sigma…
Class II FDA enforcement event 79815. Recall start date: . Status in the FDA record: Terminated. Termination date: .
Reason for recall, in the words of FDA
Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at 400ml/hr and 800ml/hr and verification of upstream occlusion detection at 800ml/hr. Improper detection of air-in-line or upstream occlusion may result in serious…
Class II, in the words of FDA
a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote
Baxter Healthcare Corporation started this device recall on 3 Apr 2018. The firm address in the FDA record is in Medina, NY. FDA put the recall in Class II. FDA event 79815 has 5 product records. The first FDA enforcement report date is 16 May 2018. The record gives the status Terminated, with the termination date 14 Feb 2020.
It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.
FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .
Distribution
In the words of FDA.
US Nationwide & Puerto Rico
Recalling firm
Baxter Healthcare Corporation, Medina, NY.
Recalling firm: The firm that the FDA record names for a recall.
FDA records how the recall started: Voluntary: Firm initiated.
Products
5 product records in the event.
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1663-2018 | Baxter Sigma Spectrum Infusion Pump with master Drug Library (Version 6) 35700BAX Product Usage: The Baxter Sigma Spectrum Infusion Pump with master Drug Library is intended to be used for the controlled administration of fluids. These fluids may includes pharmaceutical drugs, blood, blood products and mixture if… | 444 units in total | Class II | Terminated | 16 May 2018 |
| Z-1664-2018 | Baxter Sigma Spectrum Infusion Pump with master Drug Library (Version 6) 35700BAXL Product Usage: The Baxter Sigma Spectrum Infusion Pump with master Drug Library is intended to be used for the controlled administration of fluids. These fluids may includes pharmaceutical drugs, blood, blood products and mixture if… | 444 units in total | Class II | Terminated | 16 May 2018 |
| Z-1665-2018 | Baxter Sigma Spectrum Infusion Pump with Master Drug Library (Version 6) 35700BAXR Product Usage: The Baxter Sigma Spectrum Infusion Pump with master Drug Library is intended to be used for the controlled administration of fluids. These fluids may includes pharmaceutical drugs, blood, blood products and mixture if… | 444 units in total | Class II | Terminated | 16 May 2018 |
| Z-1666-2018 | Baxter Sigma Spectrum Infusion Pump with Master Drug Library (Version 6) 35700BAXRP Product Usage: The Baxter Sigma Spectrum Infusion Pump with master Drug Library is intended to be used for the controlled administration of fluids. These fluids may includes pharmaceutical drugs, blood, blood products and mixture if… | 444 units in total | Class II | Terminated | 16 May 2018 |
| Z-1667-2018 | Baxter Sigma Spectrum Infusion Pump with Master Drug Library (Version 6) 35700BAXRE Product Usage: The Baxter Sigma Spectrum Infusion Pump with master Drug Library is intended to be used for the controlled administration of fluids. These fluids may includes pharmaceutical drugs, blood, blood products and mixture if… | 444 units in total | Class II | Terminated | 16 May 2018 |
openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports.
Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.
Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.
Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.
Class II: The middle of the three FDA recall classes.
The event in context
- Device events that started in 2018
- 1,074
- Share of these events in Class II
- 92%
- Product records in this event
- 5Median of the firm: 1
- Events of the firm on the register
- 193
Other events of the firm
Official channels
- FDA enforcement report Z-1663-2018The FDA record of the first product of this event.
- FDA: recalls, market withdrawals and safety alertsThe recall notices on the FDA site.
- FDA MedWatchThe FDA safety reporting program for drugs and medical devices.
Follow by RSS:Food, drug and device recallsFDA Class I recalls
Cite this page
Recall Register. FDA enforcement event 79815. https://recallregister.com/fda/ii/79815/. Source: openFDA enforcement reports. Data through .
Sources
- Source
- openFDA device enforcement reports
- Licence
- CC0 1.0
- Export date
- 28 Sep 2026
- Newest FDA report date on the register
- 23 Sep 2026
FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.
Questions
- Where was the product distributed?
- The FDA record gives this distribution:
US Nationwide & Puerto Rico
- What quantity of product is in FDA event 79815?
- The event has 5 product records. The first record gives this quantity:
444 units in total
- How did the recall start?
- The FDA record says:
Voluntary: Firm initiated
- How many device recalls of 2018 have the same class?
- 92% of the 1,074 device events that started in 2018 are in Class II.