Food, drugs, devices › Integra LifeSciences Corp.
Device recall: The humeral stems trials are packaged in a kit for Titan Modular Shoulder System under the following labeling: Catalog N
Integra LifeSciences Corp. (Plainsboro, NJ) initiated this device recall on 20 Dec 2017; FDA lists 1 product record under event 79597, classified Class II.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 93% of the 1,091 device enforcement events initiated in 2017 carry the same class.
Status: Terminated, terminated 24 Feb 2020. Initiation: voluntary: firm initiated. Integra LifeSciences Corp. has 79 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Integra LifeSciences Corp., Plainsboro, NJ.
Reason for recall
"Incidents of stem trial breakage were reported to the firm suggesting that these fractures all occurred during insertion/impaction or extraction of the humeral stem trial whiling preparing the humeral canal and/or trialing. All cases resulted in a delay in surgery, with a variance of medical intervention required."
Distribution
"Worldwide Distribution - USA (nationwide) Distribution."
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1463-2018 | The humeral stems trials are packaged in a kit for Titan Modular Shoulder System under the following labeling: Catalog Number Product Description TRL092002506 Humeral Stem Trial, 6mm TRL092002507 Humeral Stem Trial, 7mm TRL092002508 Humeral Stem Trial, 8mm TRL092002509 Humeral Stem Trial, 9mm TRL092002510 Humeral… | 1619 units | Class II | Terminated | 2 May 2018 |
Context
Other enforcement events for Integra LifeSciences Corp.
All 79 events for Integra LifeSciences Corp.
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Incidents of stem trial breakage were reported to the firm suggesting that these fractures all occurred during insertion/impaction or extraction of the humeral stem trial whiling preparing the humeral canal and/or trialing. All cases resulted in a delay in surgery, with a variance of medical intervention required."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "Worldwide Distribution - USA (nationwide) Distribution."
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 24 Feb 2020. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.