Food, drugs, devices › Aesculap Implant Systems LLC
Device recall: DS Titanium Ligation Clip, used in laparoscopic surgery
FDA enforcement event 79530 is a device recall by Aesculap Implant Systems LLC of Center Valley, PA, initiated 22 Jan 2018 and classified Class II, with 2 product records.
FDA defines Class II as a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote. 92% of the 1,074 device enforcement events initiated in 2018 carry the same class.
Status: Terminated, terminated 24 Sep 2019. Initiation: voluntary: firm initiated. Aesculap Implant Systems LLC has 35 enforcement events on this register, with a median of 1 product record per event.
Recall classification reflects the relative degree of health hazard. Text in quotation marks is FDA's. openFDA export dated 28 Sep 2026.
Recalling firm
Aesculap Implant Systems LLC, Center Valley, PA.
Reason for recall
"Added a contraindication to IFU: Do not use the DS clips for living donor nephrectomies. Added other contraindications. A publication from 2017 recommended Aesculap DS Clips for kidney transplantation with living donors. The information from this publication was not approved or validated by the firm."
Distribution
"US and Canada distribution"
Product records
| Recall number | Product | Quantity | Class | Status | Report date |
|---|---|---|---|---|---|
| Z-1290-2018 | DS Titanium Ligation Clip, used in laparoscopic surgery | 9621 | Class II | Terminated | 11 Apr 2018 |
| Z-1291-2018 | DS Titanium Ligation Clip Appliers, used in laparoscopic surgery | 270 | Class II | Terminated | 11 Apr 2018 |
Context
Other enforcement events for Aesculap Implant Systems LLC
All 35 events for Aesculap Implant Systems LLC
Questions and answers
Why was this device recalled?
FDA's record gives the reason for recall as: "Added a contraindication to IFU: Do not use the DS clips for living donor nephrectomies. Added other contraindications. A publication from 2017 recommended Aesculap DS Clips for kidney transplantation with living donors. The information from this publication was not approved or validated by the firm."
What does Class II mean?
FDA classifies recalls by relative degree of health hazard. Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.
Where was the product distributed?
FDA's distribution pattern field reads: "US and Canada distribution"
Is the recall still ongoing?
FDA lists the status as Terminated; the termination date is 24 Sep 2019. "Terminated" means FDA has determined that all reasonable efforts to remove or correct the product have been made.
Where does this data come from?
openFDA enforcement reports (food, drug and device), CC0, export dated 28 Sep 2026. Text in quotation marks is FDA's. The FDA enforcement report linked above is authoritative.
Provenance
Source: openFDA device enforcement reports (CC0), export dated 28 Sep 2026; records grouped by FDA event number. The class shown for the event is the most severe class among its records. Errors in processing are corrected at the next build; see corrections.