Recall Register

Device recall: OTTO BOCK Kenevo knee joints Model 3C60

Class II FDA enforcement event 79019. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Otto Bock Healthcare Products GmbH has identified a design issue through extended durability testing where the pylon clamping mechanism has the potential to cause damage to the pylon. When used longer than 1.45 million cycles (which represents approximately 2.3 years of use of the knee joint under heavy usage conditions), AND the pylon clamp is overtightened according to the labeled torque…

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Otto Bock Healthcare Product started this device recall on 12 Sep 2017. The firm address in the FDA record is in Vienna. FDA put the recall in Class II. FDA event 79019 has 2 product records. The first FDA enforcement report date is 7 Mar 2018. The record gives the status Terminated, with the termination date 19 Jan 2021.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

AK, AR, AZ, CA, CT, DC, GA, IL, IN, KY, LA, MD, MI, MN, NC, ND, NJ, MN, NY, OH, OK, OR, PA, PR, TX, VA, and WA AU, BE, DE, FR, GB, IT, KW, LU, NL, NO, AT, SA, SE, CH, ES, ZA, and HU

Recalling firm

Otto Bock Healthcare Product, Vienna, N/A, Austria.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

2 product records in the event.

FDA event 79019: product records
Recall numberProductQuantityClassStatusReport date
Z-0814-2018OTTO BOCK Kenevo knee joints Model 3C6075 unitsClass IITerminated7 Mar 2018
Z-0815-2018OTTO BOCK Kenevo knee joints Model 3C60=ST4Class IITerminated7 Mar 2018

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2017
1,091
Share of these events in Class II
93%
Product records in this event
2Median of the firm: 2
Events of the firm on the register
2

Other events of the firm

Other FDA enforcement events of Otto Bock Healthcare Product
InitiatedProduct (first record)TypeClassRecords
12 Aug 2015Otto Bock Healthcare Products GmbH Genium X3 Model Number: 3B5-2, Manufacturer Otto Bock Healthcare Products GmbH. Inten Event 71950DeviceClass II1

Official channels

Cite this page

Recall Register. FDA enforcement event 79019. https://recallregister.com/fda/ii/79019/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: AK, AR, AZ, CA, CT, DC, GA, IL, IN, KY, LA, MD, MI, MN, NC, ND, NJ, MN, NY, OH, OK, OR, PA, PR, TX, VA, and WA AU, BE, DE, FR, GB, IT, KW, LU, NL, NO, AT, SA, SE, CH, ES, ZA, and HU
What quantity of product is in FDA event 79019?
The event has 2 product records. The first record gives this quantity: 75 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2017 have the same class?
93% of the 1,091 device events that started in 2017 are in Class II.