Recall Register

Device recall: daVinci X EndoWrist(R) Stapler 45, REF 470298 Product Usage: The Intuitive Surgical EndoWrist Stapler 45, Stapler 45…

Class II FDA enforcement event 78908. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

Specific da Vinci Xi EndoWrist Stapler 45 instruments are potentially impacted by a variation in the manufacturing process. In affected instruments, there is a possibility that the Stapler Release Kit (SRK) Wrench interface feature accessed through release hole 2 is manufactured to an incorrect dimension, which may prevent the SRK Wrench from being able to manually unclamp the instrument.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

Intuitive Surgical, Inc. started this device recall on 19 Sep 2017. The firm address in the FDA record is in Sunnyvale, CA. FDA put the recall in Class II. FDA event 78908 has 1 product record. The first FDA enforcement report date is 31 Jan 2018. The record gives the status Terminated, with the termination date 30 Jan 2024. 311 of the 1,091 device events that started in 2017 have more product records. The median device event of 2017 has 1 product record. The openFDA file has 1,091 device events that started in 2017 and 1,232 that started in 2016.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide Distribution

Recalling firm

Intuitive Surgical, Inc., Sunnyvale, CA.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 78908: product records
Recall numberProductQuantityClassStatusReport date
Z-0404-2018daVinci X EndoWrist(R) Stapler 45, REF 470298 Product Usage: The Intuitive Surgical EndoWrist Stapler 45, Stapler 45 Reloads, and other Stapler Accessories (including the bladeless obturators) are intended to be used with the da Vinci Surgical System (Model IS4000) for resection, transection, and/or creation of…60 unitsClass IITerminated31 Jan 2018

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2017
1,091
Share of these events in Class II
93%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
105

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 78908. https://recallregister.com/fda/ii/78908/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
10 Nov 2026
Newest FDA report date on the register
4 Nov 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide Distribution
What quantity of product is in FDA event 78908?
The FDA record gives this quantity: 60 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2017 have the same class?