Recall Register

Device recall: RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The…

Class II FDA enforcement event 78900. Recall start date: . Status in the FDA record: Terminated. Termination date: .

Reason for recall, in the words of FDA

For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used for QA measurements. This could result in an incorrect dosage being provided to a patient.

Class II, in the words of FDA

a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote

RAYSEARCH LABORATORIES AB started this device recall on 17 Jul 2017. The firm address in the FDA record is in Stockholm. FDA put the recall in Class II. FDA event 78900 has 1 product record. The first FDA enforcement report date is 24 Jan 2018. The record gives the status Terminated, with the termination date 29 Aug 2018.

It is important to note that most of the time recall classifications are determined after the recalling firm provides all of the information needed to determine the risk posed to public health. The classification and posting in the FDA Enforcement Report should not be seen as an expansion or change to a firm’s voluntary public warning.

FDA publishes these words. Source: FDA, Enforcement Reports, Important Note, read on . openFDA export of .

Distribution

In the words of FDA.

US Nationwide Distribution - FL, WA, AZ, TN, MD, TX, FL, MI, Washington D.C.

Recalling firm

RAYSEARCH LABORATORIES AB, Stockholm, N/A, Sweden.

Recalling firm: The firm that the FDA record names for a recall.

FDA records how the recall started: Voluntary: Firm initiated.

Products

1 product record in the event.

FDA event 78900: product records
Recall numberProductQuantityClassStatusReport date
Z-0380-2018RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval…117 unitsClass IITerminated24 Jan 2018

openFDA, about the status: FDA does not update the status of a recall after the recall has been classified according to its level of hazard. As such, the status of a recall (open, completed, or terminated) will remain unchanged after published in the Enforcement Reports. Source: openFDA, Drug Recall Enforcement Reports, Disclaimer.

Enforcement report: The FDA publication that lists recalls. It has one record for each product of a recall.

Enforcement event: All the product records that have one FDA event number. This site shows one event on one page.

Class II: The middle of the three FDA recall classes.

The event in context

Device events that started in 2017
1,091
Share of these events in Class II
93%
Product records in this event
1Median of the firm: 1
Events of the firm on the register
57

Other events of the firm

Official channels

Cite this page

Recall Register. FDA enforcement event 78900. https://recallregister.com/fda/ii/78900/. Source: openFDA enforcement reports. Data through .

Sources

Source
openFDA device enforcement reports
Licence
CC0 1.0
Export date
28 Sep 2026
Newest FDA report date on the register
23 Sep 2026

FDA has one record for each product. This site puts the records with the same event number on one page. The class of the event is the highest class of its records. How the register is made. Report an error.

Questions

Where was the product distributed?
The FDA record gives this distribution: US Nationwide Distribution - FL, WA, AZ, TN, MD, TX, FL, MI, Washington D.C.
What quantity of product is in FDA event 78900?
The FDA record gives this quantity: 117 units
How did the recall start?
The FDA record says: Voluntary: Firm initiated
How many device recalls of 2017 have the same class?
93% of the 1,091 device events that started in 2017 are in Class II.